FDA Adverse Event Malfunction Summary report: N

MEVATRON KD-2 LINEAR ACCELERATOR

MDR report key: 902810 · Received August 24, 2007

Report

Report Number
2910081-2007-00020
Event Type
Malfunction
Date Received
August 24, 2007
Date of Event
June 28, 2007
Report Date
August 20, 2007
Manufacturer
SIEMENS MEDICAL SOLUTIONS USA, INC.
Product Code
IYE
PMA / PMN Number
K862339
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

INITIAL COMPLAINT: "WATER LEAK IN GANTRY AT TARGET FLOW SENSOR SHORTED OUT COIL OF G55-K1 RELAY FOR STEERING CURRENT REVERSAL CIRCUIT, AND RELAY REMAINED IN DE-ENERGIZED STATE. THIS CAUSED STEERING POLARITY FOR ELECTRON ENERGIES TO BE APPLIED WITH 6MV X-RAY PROGRAMMED, AND BEAM OUTPUT WAS SUBSTANTIALLY AFFECTED DURING TREATMENTS. NUMBER (#2) TREATMENT TIME INTERLOCK WAS ASSERTED AT NEAR END OF TREATMENTS. FLATNESS & SYMMETRY INTERLOCKS DID NOT TRIP, DUE TO ORIGINAL MACHINE FWI & SWI WINDOW SETTINGS OF 8.9 & 5.0." THIS PROBLEM OCCURRED DURING PATIENT TREATMENT, BUT THE INITIAL COMPLAINT DOES NOT INDICATE HARM TO PATIENT. THIS MATTER IS CURRENTLY BEING INVESTIGATED, ALONG WITH ROOT CAUSE AND APPROPRIATE CORRECTIVE ACTION. A RISK ANALYSIS IS ALSO GOING TO BE CONDUCTED TO DETERMINE POTENTIAL HARM TO PATIENT AND OPERATOR. THIS FAULT IS CURRENTLY BEING INVESTIGATED AND AN UPDATE TO THE AGENCY WILL BE MADE ONCE FURTHER INFORMATION HAS BEEN OBTAINED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MEVATRON KD-2 LINEAR ACCELERATOR MEVATRON KD-2 IYE SIEMENS MEDICAL SOLUTIONS USA, INC. 9822685

Patients

Seq Age Sex Outcome Treatment
1 YR