ACCESS 2 IMMUNOASSAY SYSTEM
Report
- Report Number
- 2122870-2007-00199
- Event Type
- Other
- Date Received
- August 23, 2007
- Date of Event
- August 3, 2007
- Report Date
- August 23, 2007
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- NO INFORMATION
Narratives
BASED ON AVAILABLE INFO, QC WAS WITHIN SPECS AT THE BEGINNING OF THE RUN. HOWEVER, ACTUAL QC DATA WAS NOT SUPPLIED. THE CUSTOMER NOTED SOME RESULTS WITH "NO VALUE AND SYS" FLAGS, AND THE EVENT LOG INDICATED VACUUM UNDER LIMITS ERRORS. A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO THE CUSTOMER'S LAB: THE FSE INSPECTED THE INSTRUMENT AND FOUND WASTE JUG FULL. THE FSE INSTALLED AN EMPTY WASTE JUG. THE FSE INSTALLED A NEW PIPETTOR TIP AND WASH TOWER INSERT. THE FSE PERFORMED ALIGNMENTS, PRIMED LINES TO VERIFY DISPENSE AND ASPIRATION OF FLUIDS. THE FSE CONDUCTED A DIAGNOSTIC TESTING. THE FSE VERIFIED REPAIR PER ESTABLISHED PROCEDURES AND RESULTS MET PUBLISHED PERFORMANCE SPECIFICATIONS. BASED ON AVAILABLE INFO, THE CUSTOMER HAD PLACED PAPER TOWELS UNDER THE WASTE JUG WHICH CAUSED THE WASTE SENSOR TO NOT FUNCTION PROPERLY. IN A FOLLOW-UP WITH THE CUSTOMER, THEY HAVE NOT HAD ANY FURTHER ISSUES IN REGARDS TO THIS EVENT. HARDWARE ISSUES ADDRESSED BY THE FSE MAY HAVE CONTRIBUTED TO THIS EVENT. HOWEVER, A CLEAR ROOT CAUSE HAS NOT BEEN DETERMINED FOR THIS EVENT. A MALFUNCTION WILL BE ASSUMED FOR PURPOSE OF THIS REPORT.
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) REGARDING AN ERRONEOUSLY LOW PROSTATE SPECIFIC ANTIGEN (PSA) RESULT FROM A SINGLE PT SAMPLE THAT WAS GENERATED BY THE ACCESS 2 INSTRUMENT. THE INITIAL PSA RESULT WAS 0.00NG/ML. THE RESULTS WAS NOT REPORTED OUT OF THE LAB. THE ORIGINAL SAMPLE WAS RETESTED FOR PSA AND THE REPEATED RESULT WAS HIGHER: 4.59NG/ML. THE CUSTOMER DID NOT RECEIVE ANY REPORT OF PT INJURY REQUIRING MEDICAL INTERVENTION, OR CHANGE TO PT TREATMENT ATTRIBUTED OR CONNECTED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACCESS 2 IMMUNOASSAY SYSTEM | DISCRETE PHOTOMETRIC CHEMISTRY ANALYZER | JJE | BECKMAN COULTER, INC. | ACCESS 2 IMMUNOASSAY SYSTEM | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA YR |