FDA Adverse Event Other Summary report: N

ESPRIT VENTILATOR

MDR report key: 902589 · Received August 24, 2007

Report

Report Number
2031642-2007-00137
Event Type
Other
Date Received
August 24, 2007
Date of Event
August 2, 2007
Report Date
August 2, 2007
Manufacturer
RESPIRONICS CALIFORNIA, INC.
Product Code
CBK
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

NA

Description of Event or Problem · 1

WHILE SERVICING THE VENTILATOR, THE RESPIRONICS SERVICE TECHNICIAN NOTED A DIAGNOSTIC CODE IN LOG HISTORY INDICATING OCCLUSION SVO (SAFETY VALVE OPEN). THE CUSTOMER DID NOT REPORT THE OCCURRENCE DURING ANY PREVIOUS USAGE. THE RESPIRONICS SERVICE TECHNICIAN PERFORMED EXTENDED SELF TESTING (EST) WHICH FAILED DUE TO THE HEATED FILTER BACK PRESSURE BEING OUT OF RANGE. THE SPECIFICATION IS 5-15 CM H2O. THE SERVICE TECHNICIAN REPORTED THE BACK PRESSURE TEST OF THE HEATED EXHALATION FILTER MEASURED 141 CM H2O PRESSURE, INDICATING A GROSS OCCLUSION. AS SPECIFIED, A DETECTED OCCLUSION WILL CAUSE THE VENTILATOR TO ALARM AND OPEN THE SAFETY VALVE UNTIL THE OCCLUSION IS RESOLVED. THE SERVICE TECHNICIAN REPLACED THE REUSABLE EXPIRATORY FILTER TO CORRECT THE PROBLEM. EST WAS REPEATED AND PASSED TO OPERATING SPECIFICATIONS. USER MAINTENANCE CONTRIBUTED TO THIS EVENT DUE TO AN OCCLUDED FILTER. FILTER REPLACEMENT MUST BE PERFORMED AT MFR RECOMMENDED INTERVALS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ESPRIT VENTILATOR VENTILATOR, CONTINUOUS CBK RESPIRONICS CALIFORNIA, INC. V1000 NA

Patients

Seq Age Sex Outcome Treatment
1 NA YR REUSABLE EXHALATION FILTER