FDA Adverse Event Other Summary report: N

BOSTON-SCIENTIFIC

MDR report key: 902263 · Received August 20, 2007

Report

Report Number
MW5003509
Event Type
Other
Date Received
August 20, 2007
Date of Event
June 26, 2007
Report Date
August 20, 2007
Manufacturer
BOSTON SCIENTIFIC
Product Code
MNB
Report Source
Voluntary report
Reporter Location
MI, US
Reporter Occupation
NURSE

Narratives

Description of Event or Problem · 1

PATIENT DEVELOPED SECOND DEGREE VAGINAL BURN POST-HYDROTHERMAL ABLATION DUE TO EQUIPMENT POWER FAILURE DURING PROCEDURE. THE MACHINE HAD BEEN PLUGGED INTO A POWER STRIP ON THE MONITOR TOWER RATHER THAN INTO A WALL SOCKET. THIS CAUSED THE MACHINE TO POWER DOWN DURING THE BEGINNING OF THE "HEAT UP" PORTION OF THE PROCEDURE. MACHINES WERE SEPARATED AT THAT TIME AND RESTARTED PROMPTLY. THE POWER DOWN INCIDENT CAUSED HOT FLUID TO LEAK OUT OF THE MACHINE INTO THE PT'S VAGINA CAUSING THE ABOVE DESCRIBED BURNS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BOSTON-SCIENTIFIC HYDRO-THERMABLATION UNIT MNB BOSTON SCIENTIFIC

Patients

Seq Age Sex Outcome Treatment
1 32 YR Other