PERFORMER CPB SUPPORT SYSTEM
Report
- Report Number
- 3003793891-2007-00002
- Event Type
- Injury
- Date Received
- August 18, 2007
- Date of Event
- July 20, 2007
- Report Date
- July 20, 2007
- Manufacturer
- RAND S.R.L.
- Product Code
- DTQ
- PMA / PMN Number
- K052555
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- CCP
Narratives
CONCLUSION: THE DEVICE PERFORMED APPROPRIATELY, AS PROGRAMMED. IMPROPER POSITIONING OF THE BUBBLE DETECTOR AND HAVING IT SET TO STOP THE PUMP FOR VENOUS AIR RESULTED IN THE CASCADE OF EVENTS DESCRIBED IN THIS REPORT, WHICH ULTIMATELY RESULTED IN THE STOPPAGE OF FLOW FOR TWO MINUTES. THE BUBBLE DETECTOR MEANT TO STOP THE PUMP FOR ARTERIAL AIR, WHICH IS A MORE SERIOUS SITUATION. THE IFU PROVIDES EXPLICIT INSTRUCTIONS ON THE PROPER PLACEMENT OF COMPONENTS WITHIN THE CIRCUIT TO ENSURE THE SAFETY OF THE PATIENT DURING BYPASS. MEDTRONIC PRODUCT MANAGEMENT EXPLAINED THE ISSUE WITH THE HOSPITAL'S SURGICAL STAFF, WHO HAVE SINCE REPOSITIONED THE ABD TO THE ARTERIAL LINE, AND HAVE SET THE ALARM TO COAST INSTEAD OF PUMP STOP. TO DATE, THERE HAVE BEEN NO SIMILAR REPORTED OCCURRENCES.
MEDTRONIC RECEIVED INFO THAT THIS BYPASS CIRCUIT HAD BEEN SET UP IN AN OFF-LABEL CONFIGURATION, WHERE THE AIR BUBBLE DETECTOR/FLOW PROBE HAD BEEN POSITIONED ON THE VENOUS OUTLET OF THE CARDIOTOMY RESERVOIR INSTEAD OF ON THE ARTERIAL SIDE OF THE CIRCUIT. DURING THE CASE, A LOW BLOOD LEVEL ALARM OCCURRED WHICH SENT THE PUMP INTO COAST MODE (AS PROGRAMMED). DUE TO THE POSITION OF THE ABD/FLOW PROBE, AIR WAS DETECTED IN THE CIRCUIT, AND AN AIR BUBBLE ALARM OCCURRED. THIS ALARM CAUSED THE PUMP TO STOP (AS PROGRAMMED). THE PUMP FUNCTIONED NORMALLY, AS IT WAS CONFIGURED. THE PERFUSIONIST ATTEMPTED TO RE-ESTABLISH FLOW, BUT THE PARAMETERS PROGRAMMED WITHIN THE PUMP WOULD NOT ALLOW THE ARTERIAL CLAMP TO OPEN. WITH ASSISTANCE FROM MEDTRONIC PRODUCT MANAGEMENT, FLOW WAS RE-ESTABLISHED. IT WAS REPORTED THAT FLOW HAD STOPPED DURING THE CASE FOR A DURATION OF TWO-MINUTES. IT IS NOT KNOWN AT THIS TIME IF THE PATIENT SUFFERED PERMANENT INJURY DUE TO THE FLOW STOPPAGE DURING THE PROCEDURE. MEDTRONIC HAS REQUESTED DETAILS ON THE PATIENT'S STATUS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PERFORMER CPB SUPPORT SYSTEM | DTQ | RAND S.R.L. | R5100052 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | YR | Required Intervention |