FDA Adverse Event Malfunction Summary report: N

QUANTUM MAVERICK MONORAIL

MDR report key: 900448 · Received October 20, 2006

Report

Report Number
6000093-2006-02143
Event Type
Malfunction
Date Received
October 20, 2006
Date of Event
August 1, 2006
Report Date
August 1, 2006
Manufacturer
BOSTON SCIENTIFIC
Product Code
LQX
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

RETURNED PRODUCT ANALYSIS VERIFIED THE DIFFICULTY STATED IN THE COMPLAINT. PRODUCT ANALYSIS ALSO REVEALED A SHAFT KINK. THE BALLOON CATHETER WITH THE BALLOON IN A DEFLATED STATE WAS RECEIVED AND SHAFT KINK WAS OBSERVED. VISUAL AND MICROSCOPIC EXAMINATION OF THE BALLOON MATERIAL REVEALED A TEAR IN THE BALLOON WALL LOCATED 1.7 CENTIMETERS PROXIMALLY FROM THE DISTAL TIP. THE CATHETER ALSO EXHIBITED A KINK IN THE SHAFT LOCATED 51.8 CENTIMETERS DISTALLY FROM THE EDGE OF THE STRAIN RELIEF. VISUAL AND MICROSCOPIC EXAMINATION OF THE AREA SURROUNDING THE TEAR DID NOT REVEAL ANY IRREGULARITIES IN THE BALLOON MATERIAL WHICH WOULD HAVE CONTRIBUTED TO THE FORMATION OF THE TEAR. THE MFG RECORDS FOR TOP ASSEMBLY BATCH 7511497 HAVE BEEN REVIEWED AND NO ISSUES OR DISCREPANCIES WERE FOUND. THIS RECORDS REVIEW CONFIRMS THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY, AND PERFORMANCE SPECIFICATIONS. THE ROOT CAUSE OF THE INFLATION DIFFICULTIES AND SHAFT KINK COULD NOT BE DETERMINED.

Description of Event or Problem · 1

DETERMINED TO BE REPORTABLE BASED ON ANALYSIS COMPLETED ON 9/20/06. IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION PROCEDURE, INFLATION DIFFICULTIES WERE ENCOUNTERED. THE 90% STENOSED AND CALCIFIED LESION WAS LOCATED IN THE NON-TORTUOUS MID LEFT ANTERIOR DESCENDING (LAD) ARTERY. THE QUANTUM MAVERICK BALLOON EXPERIENCE REDUCED PRESSURE DURING INFLATION. THE NUMBER OF INFLATIONS AND TO WHAT ATMS IS UNK. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PT INJURIES OR COMPLICATIONS WERE NOTED. PT STATUS IS REPORTED AS "GOOD".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 QUANTUM MAVERICK MONORAIL PTCA CATHETERS LQX BOSTON SCIENTIFIC 15MM X 3.25MM 7511497

Patients

Seq Age Sex Outcome Treatment
1 YR