CONNECTOR LUER LOCK C45J
Report
- Report Number
- 3003152976-2019-00620
- Event Type
- Malfunction
- Date Received
- September 11, 2019
- Date of Event
- August 21, 2019
- Report Date
- November 14, 2019
- Manufacturer
- BECTON DICKINSON, S.A.
- Product Code
- LHI
- PMA / PMN Number
- N/A
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
H.6. INVESTIGATION: ONE SAMPLE WAS RETURNED TO OUR QUALITY TEAM FOR INVESTIGATION. UPON VISUALLY INSPECTING THE PRODUCT, NO DEFECTS WERE IDENTIFIED. A FLOW TEST WAS PERFORMED USING THE SAMPLE C45 CONNECTOR WITH A SYRINGE AND N35 INJECTOR, NO OCCLUSIONS OR FLOW ISSUES WERE OBSERVED. A DEVICE HISTORY RECORD REVIEW FOUND NO NON-CONFORMANCES ASSOCIATED WITH THIS ISSUE DURING PRODUCTION OF BATCH 1810107. PRODUCT UNDERGOES INSPECTIONS THROUGHOUT THE MANUFACTURING PROCESS TO ENSURE THE QUALITY AND FUNCTIONALITY OF THE DEVICE. TESTING WAS REVIEWED FOR THE REPORTED LOT AND ALL RESULTS WERE FOUND TO BE WITHIN SPECIFICATION. BASED ON THE AVAILABLE INFORMATION WE ARE NOT ABLE TO IDENTIFY A ROOT CAUSE RELATED TO THE MANUFACTURING PROCESS AT THIS TIME. H3 OTHER TEXT : SEE H.10.
IT WAS REPORTED THAT FLOWS ISSUES OCCURRED BEFORE USE WITH A CONNECTOR LUER LOCK C45J. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "DRUG DIDN'T FLOW. AFTER REPLACED BY NEW PRODUCT, THE ISSUE WAS RESOLVED."
A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT FLOWS ISSUES OCCURRED BEFORE USE WITH A CONNECTOR LUER LOCK C45J. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "DRUG DIDN'T FLOW. AFTER REPLACED BY NEW PRODUCT, THE ISSUE WAS RESOLVED."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 794522 | CONNECTOR LUER LOCK C45J | CONNECTOR | LHI | BECTON DICKINSON, S.A. | 1810107 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |