FDA Adverse Event Malfunction Summary report: N

CEILING TUBE MOUNT

MDR report key: 898294 · Received July 2, 2007

Report

Report Number
898294
Event Type
Malfunction
Date Received
July 2, 2007
Date of Event
June 20, 2007
Report Date
July 2, 2007
Manufacturer
SHIMADZU MEDICAL SYSTEMS
Product Code
IYB
Product Problem
Yes
Report Source
User Facility report
Reporter Location
AR, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

TUBE STAND DOES NOT EXTEND FOR PROCEDURE THAT REQUIRES PT TO STAND (DOES NOT LOWER ENOUGH). TECHS NOT AWARE THAT MANUALLY RAISING THE TUBE DOES NOT ALLOW THE COUNTER BALANCE TO RETRACT PROPERLY. SERVICE CALLED AND PARTS WERE ORDERED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CEILING TUBE MOUNT X-RAY SYSTEM, TUBE MOUNT, CEILING IYB SHIMADZU MEDICAL SYSTEMS CH30GH *

Patients

Seq Age Sex Outcome Treatment
1 *