FDA Adverse Event Malfunction Summary report: N

BD SAFETYGLIDE INSULIN SYRINGE WITH ATTACHED NEEDLE

MDR report key: 8982514 · Received September 10, 2019

Report

Report Number
1213809-2019-00916
Event Type
Malfunction
Date Received
September 10, 2019
Date of Event
August 16, 2019
Report Date
October 22, 2019
Manufacturer
BECTON DICKINSON MEDICAL SYSTEMS
Product Code
FMF
UDI-DI
30382903059035
PMA / PMN Number
SEE H.10
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: ONE PHOTO AND ONE LOOSE 1ML S/T SYRINGE WITH NEEDLE FROM SAFETYGLIDE COMBO PACKAGE WERE RECEIVED, ALONG WITH PACKAGE FROM BATCH#: 8355914 (P/N: 305903). THE SAMPLE WAS VISUALLY EVALUATED. THE SYRINGE WAS OBSERVED TO HAVE SEVERE DAMAGE DENTING THE BARREL WALL AT AN ANGLE BETWEEN APPROXIMATELY 0.3ML AND THE ZERO LINE MARKINGS. THE DAMAGE RESULTED IN REMOVING SOME OF THE SCALE MARKINGS AS WELL. THE DAMAGE OBSERVED IS REJECTABLE PER PRODUCT SPECIFICATION. POTENTIAL ROOT CAUSE FOR THE DAMAGED BARREL DEFECT IS ASSOCIATED WITH THE ASSEMBLY PROCESS. A DEVICE HISTORY RECORD REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION OF THE PROVIDED LOT NUMBER THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT. H3 OTHER TEXT : SEE H.10.

Description of Event or Problem · 0

IT WAS REPORTED THAT BEFORE USE OF THE BD SAFETYGLIDE¿ INSULIN SYRINGE WITH ATTACHED NEEDLE THE SYRINGE SHAPE IS DISTORTED AND TWISTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: SYRINGE SHAPE DISTORTION AND TWISTED.

Additional Manufacturer Narrative · 1

PMA/510(K)#: K980580 (SYRINGE) K951254 (NEEDLE). A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION AND/OR DEVICE HISTORY REVIEW, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT BEFORE USE OF THE BD SAFETYGLIDE¿ INSULIN SYRINGE WITH ATTACHED NEEDLE THE SYRINGE SHAPE IS DISTORTED AND TWISTED. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: SYRINGE SHAPE DISTORTION AND TWISTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
780589 BD SAFETYGLIDE INSULIN SYRINGE WITH ATTACHED NEEDLE PISTON SYRINGE FMF BECTON DICKINSON MEDICAL SYSTEMS 8355914 30382903059035

Patients

Seq Age Sex Outcome Treatment
1 Other