FDA Adverse Event Malfunction Summary report: N

CELL-DYN RUBY ANALYZER

MDR report key: 8982180 · Received September 10, 2019

Report

Report Number
2919069-2019-00102
Event Type
Malfunction
Date Received
September 10, 2019
Date of Event
August 20, 2019
Report Date
February 27, 2020
Manufacturer
ABBOTT DIAGNOSTICS DIVISION
Product Code
GKZ
UDI-DI
00380740017170
PMA / PMN Number
K061667
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ID
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THE FIELD SERVICE REPRESENTATIVE (FSR) OBSERVED THE CELL-DYN RUBY POWER SUPPLY WAS DAMAGED AND BURNING. SMOKE DAMAGE WAS LIMITED TO THE POWER SUPPLY AND WAS THE ONLY PART DAMAGED, NO OTHER COMPONENTS WERE AFFECTED. REPLACING THE POWER SUPPLY RESOLVED THE ISSUE. DUE TO THE DAMAGE OBSERVED IN THE POWER SUPPLY, IT COULD NOT BE DETERMINED WHERE THE TRACK DAMAGE STARTED IN THE PART. A REVIEW OF THE SERVICE HISTORY FOR THE CELL-DYN RUBY REVEALED NO ADDITIONAL ISSUES WERE REPORTED POST REPLACING OF THE POWER SUPPLY. A REVIEW OF TRACKING AND TRENDING DID NOT IDENTIFY ANY TRENDS FOR THE CELL-DYN RUBY. LABELING WAS REVIEWED AND CONTAINS ADEQUATE INFORMATION ON HAZARD, SAFETY, HOW TO MINIMIZE POTENTIAL HARM TO PERSONNEL, AND PRECAUTIONS, SUCH AS WEARING PERSONAL PROTECTIVE EQUIPMENT. BASED ON THE INFORMATION WITHIN THE COMPLAINT RECORD NO PRODUCT DEFICIENCY WAS IDENTIFIED FOR THE CELL-DYN RUBY, SN (B)(4).

Additional Manufacturer Narrative · 1

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. ALL AVAILABLE PATIENT INFORMATION WAS INCLUDED. ADDITIONAL PATIENT DETAILS ARE NOT AVAILABLE.

Description of Event or Problem · 1

THE ABBOTT FIELD SERVICE REPRESENTATIVE (FSR) REPORTED THE POWER SUPPLY WAS BURNING UPON A CELL-DYN RUBY INSTALLATION. SMOKE DAMAGE WAS LIMITED TO THE POWER SUPPLY. NO REPORTED IMPACT TO PATIENT MANAGEMENT OR USER SAFETY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
778468 CELL-DYN RUBY ANALYZER AUTOMATED HEMATOLOGY ANALYZER GKZ ABBOTT DIAGNOSTICS DIVISION 08H67-01 00380740017170

Patients

Seq Age Sex Outcome Treatment
1