FDA Adverse Event Other Summary report: N

MEDCARE LIFT WITH SCALE

MDR report key: 898131 · Received August 14, 2007

Report

Report Number
2135150-2007-00007
Event Type
Other
Date Received
August 14, 2007
Date of Event
July 17, 2007
Manufacturer
MEDCARE PRODUCTS LLC
Product Code
FSA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

IN CONVERSATION WITH MAINTENANCE, THEY INDICATED THAT THERE WERE VISIBLE SIGNS OF WEAR AND TEAR ON THE HANGAR BARS BEFORE THE INCIDENT OCCURRED. THESE HANGAR BARS WERE OVER 10 YEARS OLD, AND HAD NEVER BEEN REPLACED BY THE FACILITY. THIS MACHINE WAS NOT MANUFACTURED BY US, MEDCARE PRODUCTS, INC. IT WAS MANUFACTURED AND SOLD BY AN UNRELATED COMPANY, MEDCARE PRODUCTS, LLC. MEDCARE PRODUCTS, INC PURCHASED THE ASSETS (NOT LIABILITIES) OF MEDCARE PRODUCTS, LLC IN MARCH OF 1998.

Description of Event or Problem · 1

DURING TRANSFER FROM BATHROOM TO BED, RESIDENT WAS BEING TRANSFERRED WITH A LIFT WHEN THE YOLK HANGAR BAR THAT IS ATTACHED TO THE LOAD CELL BROKE. ONE HOOK BENT AND BROKE OFF. THE RESIDENT FELL TO THE FLOOR AND HIT THEIR HEAD. THEY WERE TAKEN TO THE HOSPITAL EMERGENCY ROOM. NO FRACTURES OR OTHER INJURIES NOTED OTHER THAN A BUMP ON THEIR HEAD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MEDCARE LIFT WITH SCALE PATIENT LIFT FSA MEDCARE PRODUCTS LLC 400003

Patients

Seq Age Sex Outcome Treatment
1 YR Other