FDA Adverse Event Malfunction Summary report: N

DA VINCI XI

MDR report key: 8978575 · Received September 10, 2019

Report

Report Number
2955842-2019-10692
Event Type
Malfunction
Date Received
September 10, 2019
Date of Event
August 19, 2019
Report Date
August 19, 2019
Manufacturer
INTUITIVE SURGICAL, INC
Product Code
NAY
PMA / PMN Number
K131861
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INTUITIVE SURGICAL, INC. (ISI) RECEIVED THE PARTS INVOLVED WITH THIS COMPLAINT AND COMPLETED THE DEVICE EVALUATION. THE REPORTED COMPLAINT WAS REPLICATED BASED ON THE FIELD EVALUATION. THE VIDEO SLICE (VSL) WAS INSTALLED AND TESTED IN THE PRINTED CIRCUIT BOARD (PCA) TEST SYSTEM. THE SYSTEM START UP FAILED WITH ERROR 307. THE VSL WAS REMOVED FROM THE SYSTEM AND POWERED UP AT THE TEST BENCH. IT CAME UP WITH NO HEARTBEAT AND ALL OF THE POWER REGULATORS WERE OFF, INDICATING A POWER PROBLEM WITH THE BOARD. BASED ON THE INFORMATION PROVIDED AT THIS TIME, THIS COMPLAINT IS BEING REPORTED BECAUSE DA VINCI SYSTEM UNAVAILABILITY AFTER THE START OF A SURGICAL PROCEDURE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI-ASSISTED SACROCOLPOPEXY SURGICAL PROCEDURE, THE SYSTEM HAD NON RECOVERABLE ERROR 307 ERROR POINTING TO THE VIDEO SLICE PROCESSING (VSL2). THE CUSTOMER HAD TO USE THE INSTRUMENT RELEASE TOOL KIT (IRT) TO REMOVE INSTRUMENTS THAT WERE STUCK ON TISSUE. THE SYSTEM WAS RESTARTED, AND THE ERROR RETURNED. TWO MORE HARD RESTARTS OF THE VISION SIDE CONSOLE (VSC) WERE PERFORMED WITH NO SUCCESS. THE PROCEDURE WAS CONVERTED TO A LAPAROSCOPIC (LAP) PROCEDURE WITH NO REPORTED INJURY. ON 8/29/2019, INTUITIVE SURGICAL INC. (ISI) CONTACTED THE OPERATING ROOM NURSE WHO CONFIRMED THAT THE ISSUE OCCURRED DURING PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO REPORT OF ANY PATIENT HARM, ADVERSE OUTCOME OR INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
772660 DA VINCI XI VISION SIDE SYSTEM, RECONDITIONED NAY INTUITIVE SURGICAL, INC 380721-05

Patients

Seq Age Sex Outcome Treatment
1