TECNIS SYMFONY TORIC
Report
- Report Number
- 9614546-2019-00843
- Event Type
- Injury
- Date Received
- September 9, 2019
- Date of Event
- June 22, 2017
- Report Date
- November 10, 2020
- Manufacturer
- JOHNSON & JOHNSON SURGICAL VISION, INC.
- Product Code
- MFK
- UDI-DI
- 05050474605176
- PMA / PMN Number
- P980040
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- PHYSICIAN
Narratives
THIS CORRECTION MDR IS BEING SUBMITTED TO ADDRESS INACCURATE INFORMATION IN SECTION F ON THE XML VERSION OF THE PREVIOUS SUBMISSION CAUSED BY A SOFTWARE GLITCH IN THE TRANSMISSION PROCESS. THE INTERNAL NON-CONFORMANCE NUMBERS FOR THIS SOFTWARE ISSUE ARE NR-(B)(4) AND CAPA-010215.
(B)(4). DEVICE EVALUATION: PRODUCT TESTING COULD NOT BE PERFORMED BECAUSE THE PRODUCT WAS NOT RETURNED. THE COMPLAINT CANNOT BE CONFIRMED. MANUFACTURING RECORD REVIEW: THE MANUFACTURING RECORDS FOR THE INTRAOCULAR LENS WERE REVIEWED. THE PRODUCT WAS MANUFACTURED ACCORDING TO SPECIFICATIONS. THE SEARCH REVEALED THAT NO SIMILAR COMPLAINT FOR THIS PRODUCTION ORDER NUMBER HAS BEEN RECEIVED. CONCLUSION: AS A RESULT OF THE INVESTIGATION THERE IS NO INDICATION OF A PRODUCT QUALITY DEFICIENCY. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC. HAS BEEN SUBMITTED.
IT WAS REPORTED THAT THE PATIENT HAD DECREASED VISION, GLARE, HAD SPIDER WEB IN VISION AND A SLIGHTLY DECENTERED INTRAOCULAR LENS (IOL) LENS. A VITRECTOMY AND A LIMBAL RELAXING INCISION (LRI) WERE REQUIRED. THERE WERE NO OTHER PATIENT INJURY. THE SURGERY WAS COMPLETED SUCCESSFULLY USING A NON-JOHNSON & JOHNSON REPLACEMENT LENS. NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 770385 | TECNIS SYMFONY TORIC | MULTIFOCAL IOLS | MFK | JOHNSON & JOHNSON SURGICAL VISION, INC. | ZXT150 | 05050474605176 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 67 YR | Required Intervention |