SOMATOM EMOTION 16 (2007)
Report
- Report Number
- 3004977335-2019-95923
- Event Type
- Injury
- Date Received
- September 9, 2019
- Date of Event
- August 28, 2019
- Report Date
- August 28, 2019
- Manufacturer
- SIEMENS HEALTHCARE GMBH
- Product Code
- JAK
- PMA / PMN Number
- K050297
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER
Narratives
(B)(6). SIEMEN'S INVESTIGATED THE REPORTED INCIDENT. THE ROOT CAUSES ARE IDENTIFIED AS FAILURE OF THE CUSTOMER TO REPAIR THE DAMAGED TABLE BEFORE USE DURING PATIENT TREATMENT AND NOT PROPERLY FIXATING THE PATIENT'S ARM DURING TREATMENT. THE REPORTED DEVICE DID NOT MALFUNCTION. A SUPPLEMENTAL REPORT WILL BE SUBMITTED IF ADDITIONAL REPORTABLE INFORMATION IS RECEIVED.
IT WAS REPORTED TO SIEMENS THAT DURING USE OF THE SOMATOM EMOTION 16 (2007) A MODERATE PATIENT INJURY OCCURRED. THE PATIENT'S FINGER WAS CAUGHT BETWEEN THE SYSTEM TABLE AND GANTRY COVER AND THE PATIENT'S SKIN WAS LACERATED BY THE SHARP EDGES OF THE PREVIOUSLY DAMAGED TABLE COVER. THE WOUND WAS TREATED DURING A BRIEF MEDICAL INTERVENTION TO STOP THE BLEEDING. THERE IS NO PERMANENT INJURY EXPECTED DUE TO THE EVENT. THE TABLE COVER WAS REPORTEDLY DAMAGED BEFORE THE REPORTED EVENT. THE EVENT IS NOT ASSOCIATED WITH AN IDENTIFIED DEVICE MALFUNCTION. THE SOURCE OF THE TABLE DAMAGE WAS NOT REPORTED BY THE CUSTOMER, NOR WAS SIEMENS NOTIFIED TO REPAIR IT BEFORE THE PATIENT EVENT. THE STRAP USED TO FIXATE PATIENT'S ARMS BECAME LOOSE AND THE PATIENT'S ARM SLIPPED INTO THE GAP BETWEEN THE DAMAGED TABLE AND THE GANTRY WHILE THE TABLE WAS MOVED INTO THE GANTRY. THE REPORTED EVENT OCCURRED IN (B)(6).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 769177 | SOMATOM EMOTION 16 (2007) | SYSTEM, X-RAY, TOMOGRAPHY, COMPUTED | JAK | SIEMENS HEALTHCARE GMBH | 1025253055 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |