AMIS CUP IMPACTOR TERMINAL-M10
Report
- Report Number
- 3005180920-2019-00738
- Event Type
- Malfunction
- Date Received
- September 6, 2019
- Date of Event
- August 7, 2019
- Report Date
- September 6, 2019
- Manufacturer
- MEDACTA INTERNATIONAL SA
- Product Code
- LXH
- UDI-DI
- 07630040736015
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- 003
Narratives
BATCH REVIEW PERFORMED ON 09-AUGUST-2019 LOT 1755175: (B)(4). NO ANOMALIES FOUND RELATED TO THE PROBLEM TO DATE, NO OTHER SIMILAR EVENTS HAVE BEEN REPORTED. VISUAL INSPECTION PERFORMED BY R&D PROJECT MANAGER. DURING THE ANALYSIS IT IS EVALUATED THAT THE BALINIT-C COATING OF THE TERMINAL IS REMOVED FROM THE AREA IN CONNECTION WITH THE CUP, IN ADDITION SOME SCRATCHES ARE PRESENT OF THE WHOLE BODY AND FINALLY THE THREADED PART IS DAMAGED. DUE TO THE DAMAGES INDICATED ABOVE THE TERMINAL WAS BLOCKED IN THE CUP AS ROOT CAUSE. ADDITIONAL DEVICE INVOLVED IN THE EVENT, BATCH REVIEW PERFORMED ON 09-AUGUST-2019 LOT 1810201: (B)(4), EXPIRATION DATE: 2024-03-12. NO ANOMALIES FOUND RELATED TO THE PROBLEM TO DATE, (B)(4) OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY OTHER SIMILAR REPORTED EVENT.
AFTER THE SURGEON IMPACTED THE CUP, IT WAS NOT POSSIBLE TO DISCONNECT THE TERMINAL CUP IMPACTOR. HE HAD TO EXPLANT THE CUP AND USE A NEW ONE WITH THE STRAIGHT CUP IMPACTOR. THERE WAS A 5 MINUTES DELAY
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 764959 | AMIS CUP IMPACTOR TERMINAL-M10 | SURGICAL INSTRUMENT FOR HIP SURGERY | LXH | MEDACTA INTERNATIONAL SA | 1755175 | 07630040736015 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |