AMIS CUP IMPACTOR TERMINAL-M10

MDR report key: 8969653 · Received September 6, 2019

Report

Report Number
3005180920-2019-00738
Event Type
Malfunction
Date Received
September 6, 2019
Date of Event
August 7, 2019
Report Date
September 6, 2019
Manufacturer
MEDACTA INTERNATIONAL SA
Product Code
LXH
UDI-DI
07630040736015
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

BATCH REVIEW PERFORMED ON 09-AUGUST-2019 LOT 1755175: (B)(4). NO ANOMALIES FOUND RELATED TO THE PROBLEM TO DATE, NO OTHER SIMILAR EVENTS HAVE BEEN REPORTED. VISUAL INSPECTION PERFORMED BY R&D PROJECT MANAGER. DURING THE ANALYSIS IT IS EVALUATED THAT THE BALINIT-C COATING OF THE TERMINAL IS REMOVED FROM THE AREA IN CONNECTION WITH THE CUP, IN ADDITION SOME SCRATCHES ARE PRESENT OF THE WHOLE BODY AND FINALLY THE THREADED PART IS DAMAGED. DUE TO THE DAMAGES INDICATED ABOVE THE TERMINAL WAS BLOCKED IN THE CUP AS ROOT CAUSE. ADDITIONAL DEVICE INVOLVED IN THE EVENT, BATCH REVIEW PERFORMED ON 09-AUGUST-2019 LOT 1810201: (B)(4), EXPIRATION DATE: 2024-03-12. NO ANOMALIES FOUND RELATED TO THE PROBLEM TO DATE, (B)(4) OF THE SAME LOT HAVE BEEN ALREADY SOLD WITHOUT ANY OTHER SIMILAR REPORTED EVENT.

Description of Event or Problem · 1

AFTER THE SURGEON IMPACTED THE CUP, IT WAS NOT POSSIBLE TO DISCONNECT THE TERMINAL CUP IMPACTOR. HE HAD TO EXPLANT THE CUP AND USE A NEW ONE WITH THE STRAIGHT CUP IMPACTOR. THERE WAS A 5 MINUTES DELAY

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
764959 AMIS CUP IMPACTOR TERMINAL-M10 SURGICAL INSTRUMENT FOR HIP SURGERY LXH MEDACTA INTERNATIONAL SA 1755175 07630040736015

Patients

Seq Age Sex Outcome Treatment
1 Other