FDA Adverse Event Malfunction Summary report: N

CATALYS SYSTEM

MDR report key: 8969480 · Received September 6, 2019

Report

Report Number
3005675890-2019-00074
Event Type
Malfunction
Date Received
September 6, 2019
Date of Event
August 15, 2019
Report Date
November 10, 2020
Manufacturer
JOHNSON & JOHNSON SURGICAL VISION, INC.
Product Code
OOE
UDI-DI
05050474609792
PMA / PMN Number
K121091
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

CORRECTION: IN INITIAL REPORT, ONLY THE MANUFACTURE YEAR; 2016 WAS PROVIDED. THE CORRECT DATE IS 12/8/2014. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC HAS BEEN SUBMITTED.

Additional Manufacturer Narrative · 0

THIS CORRECTION MDR IS BEING SUBMITTED TO ADDRESS INACCURATE INFORMATION IN SECTION F ON THE XML VERSION OF THE PREVIOUS SUBMISSION CAUSED BY A SOFTWARE GLITCH IN THE TRANSMISSION PROCESS. THE INTERNAL NON-CONFORMANCE NUMBERS FOR THIS SOFTWARE ISSUE ARE NR-0148919 AND CAPA-010215.

Additional Manufacturer Narrative · 1

JOHNSON & JOHNSON SURGICAL VISION, INC INVESTIGATION SUBSEQUENTLY IDENTIFIED THAT THE CATALYS SYSTEM MAY HAVE A REMOTE POTENTIAL EVENT WHERE LOSS OF SUCTION DURING LASER FIRING COULD CREATE A HAZARDOUS SITUATION THAT MAY RESULT IN SCORING OF THE POSTERIOR CORNEAL SURFACE. THEREFORE JOHNSON & JOHNSON SURGICAL VISION, INC HAS ISSUED AN ADVISORY NOTICE TO REINFORCE INSTRUCTIONS PROVIDED IN PRODUCT TRAINING AND IN THE OPERATOR¿S MANUAL REGARDING SUCTION LOSS DURING A PROCEDURE. A REVIEW OF RECORDS RELATED TO THE DEVICE INCLUDING LABELING, COMPLAINT TRENDING, AND RISK DOCUMENTATION WILL BE PERFORMED. UPON COMPLETION OF THIS REVIEW, IF THERE IS ANY FURTHER RELEVANT INFORMATION OBTAINED, A SUPPLEMENTAL MEDWATCH WILL BE FILED. ALL PERTINENT INFORMATION AVAILABLE TO JOHNSON & JOHNSON SURGICAL VISION, INC HAS BEEN SUBMITTED.

Description of Event or Problem · 1

THE SURGERY CENTER REPORTED SUCTION LOSS DURING THE CATALYS TREATMENT. SUCTION WAS LOST DURING FRAGMENTATION AND THE CATALYST PROCEDURE WAS NOT COMPLETED. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
764454 CATALYS SYSTEM CATALYS OOE JOHNSON & JOHNSON SURGICAL VISION, INC. CATALYS-U 05050474609792

Patients

Seq Age Sex Outcome Treatment
1