INRATIO
Report
- Report Number
- 2954730-2007-00366
- Event Type
- Malfunction
- Date Received
- July 13, 2007
- Date of Event
- July 24, 2007
- Report Date
- August 9, 2007
- Manufacturer
- HEMOSENSE, INC.
- Product Code
- JPA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED: DATE: (UNKNOWN) 2007; INRATIO: 5.2, LAB: 3.76, MEAN: 4.48, CONFIDENCE LIMITS: 2.5 - 6.5; DATE: (070281) 2007; INRATIO: 4.7, LAB: 1.9, MEAN: 3.3, CONFIDENCE LIMITS: 2.0 - 5.0. PER INTERNAL PROCEDURE, THE MEAN OF THE INRATIO METER AND COMPARATIVE SYSTEM INR WERE CALCULATED. FOR THE 1ST DATA SET, UNKNOWN STRIP LOT NUMBER, BOTH INRATIO AND LAB VALUES ARE WITHIN THE CONFIDENCE LIMITS FOR INR TESTING. FOR THE SECOND DATA SET, STRIP LOT NUMBER 070281, THE LAB INR IS NOT WITHIN THE CONFIDENCE LIMITS. THE RESULTS ARE CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. THEREFORE, FURTHER TESTING OF STRIP LOT 070281 IS REQUIRED AT THIS TIME.
CALLER ALLEGES INACCURACY WITH INRATIO. RESULTS AS FOLLOWS: DATE: (UNKNOWN) 2007, INRATIO: 5.2, LAB: 3.76; DATE: 2007; INRATIO: 4.7, LAB: 1.9.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INRATIO | PROTHROMBIN TIME TEST | JPA | HEMOSENSE, INC. | 070281 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | YR |