FDA Adverse Event Malfunction Summary report: N

INRATIO

MDR report key: 896703 · Received July 13, 2007

Report

Report Number
2954730-2007-00366
Event Type
Malfunction
Date Received
July 13, 2007
Date of Event
July 24, 2007
Report Date
August 9, 2007
Manufacturer
HEMOSENSE, INC.
Product Code
JPA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED: DATE: (UNKNOWN) 2007; INRATIO: 5.2, LAB: 3.76, MEAN: 4.48, CONFIDENCE LIMITS: 2.5 - 6.5; DATE: (070281) 2007; INRATIO: 4.7, LAB: 1.9, MEAN: 3.3, CONFIDENCE LIMITS: 2.0 - 5.0. PER INTERNAL PROCEDURE, THE MEAN OF THE INRATIO METER AND COMPARATIVE SYSTEM INR WERE CALCULATED. FOR THE 1ST DATA SET, UNKNOWN STRIP LOT NUMBER, BOTH INRATIO AND LAB VALUES ARE WITHIN THE CONFIDENCE LIMITS FOR INR TESTING. FOR THE SECOND DATA SET, STRIP LOT NUMBER 070281, THE LAB INR IS NOT WITHIN THE CONFIDENCE LIMITS. THE RESULTS ARE CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. THEREFORE, FURTHER TESTING OF STRIP LOT 070281 IS REQUIRED AT THIS TIME.

Description of Event or Problem · 1

CALLER ALLEGES INACCURACY WITH INRATIO. RESULTS AS FOLLOWS: DATE: (UNKNOWN) 2007, INRATIO: 5.2, LAB: 3.76; DATE: 2007; INRATIO: 4.7, LAB: 1.9.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INRATIO PROTHROMBIN TIME TEST JPA HEMOSENSE, INC. 070281

Patients

Seq Age Sex Outcome Treatment
1 YR