FDA Adverse Event Malfunction Summary report: N

SYRINGE 60ML LL TIP 1ML 2 OZ IN 1/4 OZ

MDR report key: 8965883 · Received September 5, 2019

Report

Report Number
1911916-2019-00912
Event Type
Malfunction
Date Received
September 5, 2019
Date of Event
July 1, 2019
Report Date
October 4, 2019
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
FMF
UDI-DI
30382903096535
PMA / PMN Number
K980987
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: SAMPLES FOR THIS COMPLAINT WERE SENT BY THE CUSTOMER; HOWEVER, WE ARE UNABLE TO LOCATE THEM AT THIS TIME. SHOULD THEY BE LOCATED WE WILL RE-INVESTIGATE. A DEVICE HISTORY RECORD (DHR) REVIEW WAS PERFORMED ON THE BATCH ASSOCIATED WITH THIS INVESTIGATION. THE DHR REVIEWS DID NOT REVEAL ANY DEFECTS OR ISSUES REPORTED AND NO QUALITY NOTIFICATIONS WERE ISSUED. BASED ON THE INVESTIGATION CONCLUSION AND WITHOUT A SAMPLE TO ANALYZE, THE CONDITION REPORTED BY THE CUSTOMER COULD NOT BE CONFIRMED. THEREFORE, A POTENTIAL ROOT CAUSE(S) CANNOT BE DETERMINED, AND NO CORRECTIVE OR PREVENTATIVE ACTIONS ARE PROPOSED IN THE SCOPE OF THIS COMPLAINT.

Description of Event or Problem · 0

IT WAS REPORTED THAT 22 SYRINGE 60ML LL TIP 1ML 2 OZ IN 1/4 OZ EXPERIENCED FOREIGN MATTER CONTAMINATION WHICH WAS NOTED PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: DURING THE INSPECTION OF BATCH NUMBER 8220711, SYRINGE 50 ML OF BD FROM USA, 22 SYRINGES FOUND WITH CLEARLY VISIBLE PARTICLES AND FLIEBERS. THERE WERE MANY MORE SYRINGES WITH PARTICLES BUT THEY WERE NOT BIG ENOUGH.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT 22 SYRINGE 60 ML LL TIP 1 ML 2 OZ IN 1/4 OZ EXPERIENCED FOREIGN MATTER CONTAMINATION WHICH WAS NOTED PRIOR TO USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: DURING THE INSPECTION OF BATCH NUMBER 8220711, SYRINGE 50 ML OF BD FROM USA, 22 SYRINGES FOUND WITH CLEARLY VISIBLE PARTICLES AND FIBERS. THERE WERE MANY MORE SYRINGES WITH PARTICLES BUT THEY WERE NOT BIG ENOUGH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
761884 SYRINGE 60ML LL TIP 1ML 2 OZ IN 1/4 OZ PISTON SYRINGE FMF BD MEDICAL (BD WEST) MEDICAL SURGICAL 8220711 30382903096535

Patients

Seq Age Sex Outcome Treatment
1 Other