FAST-CATH¿ TRANSSEPTAL GUIDING INTRODUCER SWARTZ¿ SL TRANSSEPTAL SERIES 63 CM LE
Report
- Report Number
- 3005334138-2019-00485
- Event Type
- Injury
- Date Received
- September 4, 2019
- Report Date
- September 4, 2019
- Manufacturer
- ST. JUDE MEDICAL
- Product Code
- DYB
- PMA / PMN Number
- K061015
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE RESULTS OF THE INVESTIGATION ARE INCONCLUSIVE SINCE THE DEVICE WAS NOT RETURNED FOR ANALYSIS. BASED ON THE INFORMATION RECEIVED, THE CAUSE OF THE REPORTED DETACHMENT COULD NOT BE CONCLUSIVELY DETERMINED. THE DEVICE HISTORY RECORD WAS REVIEWED TO ENSURE THAT EACH MANUFACTURING AND INSPECTION OPERATION WAS PERFORMED.
THE FOLLOWING WAS PUBLISHED IN CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS IN AN ARTICLE TITLED ¿RETRIEVAL OF A DETACHED TRANSSEPTAL SHEATH TIP FROM A RIGHT PULMONARY ARTERY BRANCH FOLLOWING CATHETER ABLATION" BY SCHRICKER ET AL. 2015. 86:1131. ¿DURING REMOVAL OF THE REMAINING SL1, THE TERMINAL 2-CM TIP OF THE SHEATH BROKE OFF WITHIN THE RIGHT FEMORAL VEIN AS THE SHEATH WAS BEING REMOVED FROM THE VEIN. THERE WAS NO EXCESSIVE FORCE OR PULLING OF THE SHEATH DURING REMOVAL, ALTHOUGH THERE WAS SOME RESISTANCE TO ITS REMOVAL INITIALLY. FLUOROSCOPY DEMONSTRATED THAT THE RADIOPAQUE TIP HAD EMBOLIZED AND WAS OSCILLATING IN THE UPPER RIGHT ATRIUM NEAR THE RIGHT ATRIAL¿SUPERIOR VENA CAVA JUNCTION. SHEATH TIP RETRIEVAL TOOK 90 MINUTES, WHICH INCLUDED 60 MIN OF FLUOROSCOPY. THE PATIENT EXPERIENCED NO IMMEDIATE COMPLICATIONS, AND 15 MONTHS POSTPROCEDURE HE REMAINS FREE OF BOTH PROCEDURAL RELATED COMPLICATIONS AND ARRHYTHMIA.¿ (DOI: 10.1002/CCD.25957) .
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 756498 | FAST-CATH¿ TRANSSEPTAL GUIDING INTRODUCER SWARTZ¿ SL TRANSSEPTAL SERIES 63 CM LE | INTRODUCER, CATHETER | DYB | ST. JUDE MEDICAL | 406840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR | Other| R |