FDA Adverse Event Malfunction Summary report: N

BD INSULIN SYRINGE WITH BD ULTRA-FINE NEEDLE

MDR report key: 8961065 · Received September 4, 2019

Report

Report Number
1920898-2019-00904
Event Type
Malfunction
Date Received
September 4, 2019
Date of Event
August 13, 2019
Report Date
September 20, 2019
Manufacturer
BD MEDICAL - DIABETES CARE
Product Code
FMF
PMA / PMN Number
N/A
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NU
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: CUSTOMER RETURNED PHOTOS OF SHELF CARTONS FOR 1/2CC, 6MM, 31G SYRINGES FROM LOT#: 5222579. CUSTOMER STATES THAT THE BOX IS DAMAGED AND THE LOT IS BURRED AND EXPIRED. THE ATTACHED PHOTOS WERE EXAMINED AND EXHIBITED TORN SHELF CARTONS. ALL LOT NUMBERS WERE PRINTED CLEARLY ON THE SHELF CARTONS AND THE EXPIRATION DATE PRINTED ON THE SHELF CARTON IN THE ATTACHED PHOTOS READS AS AUGUST 2020, WHICH INDICATES THAT THE PRODUCT IS NOT EXPIRED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR BATCH#: 5222579. ALL INSPECTIONS AND CHALLENGES WERE PERFORMED PER THE APPLICABLE OPERATIONS QC SPECIFICATIONS. THERE WERE ZERO NOTIFICATIONS NOTED THAT PERTAINED TO THE COMPLAINT. BD WAS ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE (TORN SHELF CARTONS). BD WAS NOT ABLE TO DUPLICATE OR CONFIRM THE CUSTOMER¿S INDICATED FAILURE (BLURRY AND EXPIRED LOT). H3 OTHER TEXT.

Description of Event or Problem · 0

IT WAS REPORTED THAT DAMAGED PACKAGING WAS FOUND BEFORE USE WITH A BD INSULIN SYRINGE WITH BD ULTRA-FINE¿ NEEDLE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER. "DAMAGED BOX, BLURRED AND EXPIRED LOT, CODE: 324917, BATCH: 5222579". 7 OCCURRENCES WERE REPORTED, BUT THE DATES/TIMES WERE NOT GIVEN.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DAMAGED PACKAGING WAS FOUND BEFORE USE WITH A BD INSULIN SYRINGE WITH BD ULTRA-FINE¿ NEEDLE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "DAMAGED BOX BLURRED AND EXPIRED LOT CODE: 324917 BATCH: 5222579" 7 OCCURRENCES WERE REPORTED, BUT THE DATES/TIMES WERE NOT GIVEN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
754429 BD INSULIN SYRINGE WITH BD ULTRA-FINE NEEDLE SYRINGE FMF BD MEDICAL - DIABETES CARE 5222579

Patients

Seq Age Sex Outcome Treatment
1 Other