INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2019-12498
- Event Type
- Injury
- Date Received
- September 3, 2019
- Date of Event
- August 3, 2019
- Report Date
- October 16, 2020
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FTR
- PMA / PMN Number
- P020056
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENT OF RUPTURE AND CYST WAS RECEIVED ON MARCH 17, 2020 WITH LOT NUMBER: 3142077. ANALYSIS OF THE RETURNED DEVICE IDENTIFIED: UNDERWEIGHT, OPENING ON POSTERIOR, RED PARTICLES IN THE SHELL. A VISUAL AND MICROSCOPIC ANALYSIS WAS PERFORMED WHICH IDENTIFIED SHARP OPENING AND CREASE FOLD. A DIMENSION MEASUREMENT IN THE SHELL WAS PERFORMED WHICH IDENTIFY THE THICKNESS WITHIN SPECIFICATION. BASE ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS: SHARP OPENING ON POSTERIOR ASSESSED AS AN UNIDENTIFIED (TEAR) OPENING. ALLERGAN DID NOT SUBMIT THIS MDR WITHIN 30 DAYS OF BECOMING AWARE. RECENT STIMULATED REPORTING RELATED TO 2011068-7/2/19-001-R HAS INCREASED COMPLAINT AND MDR VOLUME. ALLERGAN IS IMPLEMENTING A PLAN TO ADDRESS THE INCREASED VOLUMES.
ALLERGAN DID NOT SUBMIT THIS MDR WITHIN 30 DAYS OF BECOMING AWARE. RECENT STIMULATED REPORTING RELATED TO 2011068-7/2/19-001-R HAS INCREASED COMPLAINT AND MDR VOLUME. ALLERGAN IS IMPLEMENTING A PLAN TO ADDRESS THE INCREASED VOLUMES.
HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE RUPTURE AND CYST. THE DEVICE HAS BEEN EXPLANTED.
THE DEVICE WAS EXPLANTED.
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. THE REASON FOR REOPERATION: RUPTURE. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. THIS IS A KNOWN POTENTIAL ADVERSE EVENT ADDRESSED IN THE PRODUCT LABELING.
HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE RUPTURE AND CYST. THE DEVICE REMAINS IMPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 750597 | INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | ALLERGAN (COSTA RICA) | 3142077 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 35 YR | Required Intervention |