FDA Adverse Event Injury Summary report: N

INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 8958632 · Received September 3, 2019

Report

Report Number
9617229-2019-12498
Event Type
Injury
Date Received
September 3, 2019
Date of Event
August 3, 2019
Report Date
October 16, 2020
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENT OF RUPTURE AND CYST WAS RECEIVED ON MARCH 17, 2020 WITH LOT NUMBER: 3142077. ANALYSIS OF THE RETURNED DEVICE IDENTIFIED: UNDERWEIGHT, OPENING ON POSTERIOR, RED PARTICLES IN THE SHELL. A VISUAL AND MICROSCOPIC ANALYSIS WAS PERFORMED WHICH IDENTIFIED SHARP OPENING AND CREASE FOLD. A DIMENSION MEASUREMENT IN THE SHELL WAS PERFORMED WHICH IDENTIFY THE THICKNESS WITHIN SPECIFICATION. BASE ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS: SHARP OPENING ON POSTERIOR ASSESSED AS AN UNIDENTIFIED (TEAR) OPENING. ALLERGAN DID NOT SUBMIT THIS MDR WITHIN 30 DAYS OF BECOMING AWARE. RECENT STIMULATED REPORTING RELATED TO 2011068-7/2/19-001-R HAS INCREASED COMPLAINT AND MDR VOLUME. ALLERGAN IS IMPLEMENTING A PLAN TO ADDRESS THE INCREASED VOLUMES.

Additional Manufacturer Narrative · 0

ALLERGAN DID NOT SUBMIT THIS MDR WITHIN 30 DAYS OF BECOMING AWARE. RECENT STIMULATED REPORTING RELATED TO 2011068-7/2/19-001-R HAS INCREASED COMPLAINT AND MDR VOLUME. ALLERGAN IS IMPLEMENTING A PLAN TO ADDRESS THE INCREASED VOLUMES.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE RUPTURE AND CYST. THE DEVICE HAS BEEN EXPLANTED.

Description of Event or Problem · 0

THE DEVICE WAS EXPLANTED.

Additional Manufacturer Narrative · 1

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. THE REASON FOR REOPERATION: RUPTURE. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. THIS IS A KNOWN POTENTIAL ADVERSE EVENT ADDRESSED IN THE PRODUCT LABELING.

Description of Event or Problem · 1

HEALTHCARE PROFESSIONAL REPORTED RIGHT SIDE RUPTURE AND CYST. THE DEVICE REMAINS IMPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
750597 INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 3142077

Patients

Seq Age Sex Outcome Treatment
1 35 YR Required Intervention