FDA Adverse Event Malfunction Summary report: N

TWISTR ACL DISPOSABLES KIT

MDR report key: 8958076 · Received September 3, 2019

Report

Report Number
1221934-2019-58184
Event Type
Malfunction
Date Received
September 3, 2019
Date of Event
January 1, 2019
Report Date
September 3, 2019
Manufacturer
MEDOS INTERNATIONAL SàRL
Product Code
KDD
UDI-DI
10886705026173
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CI
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY: THE COMPLAINT DEVICE WAS NOT RETURNED, THEREFORE A PHYSICAL EVALUATION CANNOT BE PERFORMED. THE COMPLAINT EVENT WAS REPORTED TO HAVE OCCURRED DURING USE ON A CADAVER. A PHOTO WAS SUPPLIED AT THE TIME THE COMPLAINT WAS CREATED, THIS WAS USED TO EVALUATE THE COMPLAINT DEVICE. THE PHOTO REVEALED THE DISTAL TIP OF THE 14 INCH DRILL PIN WAS EXTENDING OUT OF THE BULLET, APPEARED TO BE BENT AT THE TIP, AND WAS BROKEN OFF COMPLETELY WHERE IT COMES OUT OF THE OTHER END OF THE BULLET. THE DRILL PIN APPEARS TO BE STUCK INSIDE THE BULLET, AS REPORTED IN THE EVENT DESCRIPTION. THIS COMPLAINT IS CONFIRMED. THE PHOTO WAS REVIEWED BY AN NPD ENGINEER AS PART OF THE EVALUATION. THE NPD ENGINEER CONCLUDED THAT THE DRILL PIN LIKELY GOT GALLED UP IN THE BULLET, AND IN TRYING TO GET THEM DECOUPLED, THE WELD BETWEEN THE BULLET AND HANDLE FAILED. IF THE DRILL PIN WAS NOT STRAIGHT, IT COULD HAVE STARTED TO GALL DURING THE DRILLING, WHICH EVENTUALLY CAUSED IT TO SEIZE TOGETHER. IT IS LIKELY THAT THE DRILL PIN BROKE INTO TWO PIECES DURING THE ATTEMPT TO REMOVE THE STUCK PIN FROM THE BULLET, AT THE SAME TIME THE HANDLE BROKE OFF THE BULLET. A MANUFACTURING RECORD EVALUATION CANNOT BE CONDUCTED AS NO LOT NUMBERS WERE PROVIDED. AT THIS POINT IN TIME, NO CORRECTIVE ACTION IS REQUIRED, AND NO FURTHER ACTION IS WARRANTED. HOWEVER, DEPUY SYNTHES MITEK WILL CONTINUE TO TRACK ANY RELATED COMPLAINTS WITHIN THIS DEVICE FAMILY AS A MEANS OF MONITORING THE EXTENT WITH WHICH THIS COMPLAINT IS OBSERVED IN THE FIELD. UDI: (B)(4). THE LOT NUMBER IS UNKNOWN. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

IT WAS REPORTED BY THE AFFILIATE VIA EMAIL THAT THE FAILURE OF THE CRUCIATE+ BEATH PIN BULLET WAS IDENTIFIED DURING ITS CORRECT USE IN A CADAVEROUS COURSE, THE BEATH PIN 14'', WHICH WAS A PART OF THE DISPOSABLE KIT GOT TRAPPED AND LOCKED INSIDE THE BULLET. WHEN TRYING TO RELEASE IT, THE BULLET WAS BROKEN AND WAS REMAINING IN 2 PARTS IMAGES ATTACHED. BUT STILL THE DRILL DID NOT COME OUT. THE RECEIVED PHOTO WAS REVIEWED BY MANUFACTURER AND IT SHOWS A RESTORE ACL INSTRUMENTS 14" DRILL PT. PIN W/EYELET STUCK INSIDE THE BULLET COMPLAINT DEVICE. IT IS CLEAR IN THE PHOTO THAT THE DRILL PIN IS ABOUT HALF THE LENGTH IT IS SUPPOSED TO BE AND ALSO APPEARS BENT AT THE TIP. THE DRILL PIN IS PART OF THE FINISHED GOODS ACL DISPOSABLE KIT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
750805 TWISTR ACL DISPOSABLES KIT LIGAMENT RECONSTRUCTION INSTRUMENT SET KDD MEDOS INTERNATIONAL SàRL UNKNOWN 10886705026173

Patients

Seq Age Sex Outcome Treatment
1