FDA Adverse Event Malfunction Summary report: N

BD LUER-LOK TIP SYRINGE

MDR report key: 8956425 · Received September 3, 2019

Report

Report Number
1911916-2019-00897
Event Type
Malfunction
Date Received
September 3, 2019
Date of Event
August 14, 2019
Report Date
September 11, 2019
Manufacturer
BD MEDICAL (BD WEST) MEDICAL SURGICAL
Product Code
FMF
UDI-DI
30382903028321
PMA / PMN Number
K980987
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ND, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION: FOUR (4) SAMPLES WERE RECEIVED FROM THE CUSTOMER FOR INVESTIGATION. THE RETURNED SAMPLES WERE VISUALLY EXAMINED, TWO (2) OF THE RETURNED SAMPLES HAD MINOR SCALE PRINTING ISSUED BUT NO SCRATCHES WERE OBSERVED IN THE BARRELS. A DEVICE HISTORY RECORD (DHR) REVIEW WAS PERFORMED ON THE BATCHES ASSOCIATED WITH THIS INVESTIGATION. THE DHR REVIEW DID NOT REVEAL ANY DEFECTS OR ISSUES REPORTED AND NO QUALITY NOTIFICATIONS WERE ISSUED. BASED ON THE INVESTIGATION CONCLUSION THE CONDITION REPORTED BY THE CUSTOMER CANNOT BE CONFIRMED.

Description of Event or Problem · 0

MATERIAL NO.: 302832 BATCH NO.: 9182467, 9149628, 9120516 IT WAS REPORTED THAT DURING USE OF THE BD LUER-LOK¿ TIP SYRINGE THERE ARE SCRATCHES ON THE BARREL. THERE ARE SCRATCHES ON THE BARREL OF THE SYRINGE.

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 9182467. MEDICAL DEVICE EXPIRATION DATE: 2024-06-30. DEVICE MANUFACTURE DATE: 2019-07-01. MEDICAL DEVICE LOT #: 9149628. MEDICAL DEVICE EXPIRATION DATE: 2024-05-31. DEVICE MANUFACTURE DATE: 2019-05-29. MEDICAL DEVICE LOT #: 9120516. MEDICAL DEVICE EXPIRATION DATE: 2024-04-30. DEVICE MANUFACTURE DATE: 2019-04-30. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

MATERIAL NO.: 302832 BATCH NO.: 9182467, 9149628, 9120516. IT WAS REPORTED THAT DURING USE OF THE BD LUER-LOK¿ TIP SYRINGE THERE ARE SCRATCHES ON THE BARREL. THERE ARE SCRATCHES ON THE BARREL OF THE SYRINGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
750640 BD LUER-LOK TIP SYRINGE PISTON SYRINGE FMF BD MEDICAL (BD WEST) MEDICAL SURGICAL SEE H.10 30382903028321

Patients

Seq Age Sex Outcome Treatment
1 Other