FDA Adverse Event Malfunction Summary report: N

APEX¿

MDR report key: 8955177 · Received September 3, 2019

Report

Report Number
1641965-2019-00014
Event Type
Malfunction
Date Received
September 3, 2019
Date of Event
August 7, 2019
Report Date
October 9, 2019
Manufacturer
B. BRAUN MEDICAL INC
Product Code
NEP
UDI-DI
04046964801068
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MEDICAL INC. INTERNAL REPORT NUMBER (B)(4). NO LOGS OR FILES WERE RETURNED FOR EVALUATION. TO DETERMINE A POTENTIAL ROOT CAUSE A COPY OF THE DATABASE OR THE PATIENT FILE IS REQUIRED AND THEREFORE THE EVENT CANNOT BE CONFIRMED. THIS IS LIKELY AN ISSUE WITH THE CUSTOMER'S SETUP OF THE DATABASE OR HOW THE PATIENT FILE (.PAT) WAS GENERATED. IF ADDITIONAL PERTINENT INFORMATION BECOMES AVAILABLE A FOLLOW-UP REPORT WILL BE FILED.

Description of Event or Problem · 0

AS PER USER FACILITY: CUSTOMER REPORTED THEY HAD ADDED THE NEW SMOFLIPID TO THEIR SCDM DATABASE EARLIER IN THE DAY BUT THE COMPOUNDER WAS NOT PUMPING THE CORRECT AMOUNT FOR THE PATIENT ORDER. THE COMPOUNDER WAS PUMPING WATER INSTEAD OF THE SMOFLIPID SOLUTION. NO PATIENT INVOLVEMENT.

Additional Manufacturer Narrative · 1

THIS REPORT HAS BEEN IDENTIFIED AS B. BRAUN MEDICAL INC. INTERNAL REPORT NUMBER (B)(4). THE INVESTIGATION IS ONGOING AT THIS TIME. A FOLLOW UP WILL BE SUBMITTED WHEN THE INVESTIGATION RESULTS BECOME AVAILABLE.

Description of Event or Problem · 1

AS PER USER FACILITY: CUSTOMER REPORTED THEY HAD ADDED THE NEW SMOFLIPID TO THEIR SCDM DATABASE EARLIER IN THE DAY BUT THE COMPOUNDER WAS NOT PUMPING THE CORRECT AMOUNT FOR THE PATIENT ORDER. THE COMPOUNDER WAS PUMPING WATER INSTEAD OF THE SMOFLIPID SOLUTION. NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
749563 APEX¿ COMPOUNDER NEP B. BRAUN MEDICAL INC 601224 04046964801068

Patients

Seq Age Sex Outcome Treatment
1