FDA Adverse Event
Malfunction
Summary report: N
PLATO BPS V. 13.2
MDR report key: 89529
·
Received May 6, 1997
Report
- Report Number
- 1121753-1997-00015
- Event Type
- Malfunction
- Date Received
- May 6, 1997
- Report Date
- April 17, 1997
- Manufacturer
- NUCLETRON CORP.
- Product Code
- IWB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- UNKNOWN
Narratives
Description of Event or Problem · 1
FOR SOFTWARE V. 13.2, IF AN EXISTING PLAN IS SELECTED, THEN READ IN A AS A CUSTOM FILE WITH DIFFERENT STEP SIZE, THE PROGRAM ALLOWS ONE TO WRITE A PROGRAM WITH A DIFFERENT STEP SIZE THAN THE STEP SIZE ORIGINALLY PROGRAMMED. THE JOURNAL FILE PRINTOUT DISPLAYS THE MESSAGES: (1) "INACCURATE DOSE CALCULATION MAY RESULT" AND (2) PROGRAM CARD IS DISABLED". HOWEVER THE PROGRAM CARD IS NOT DISABLED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PLATO BPS V. 13.2 | RADIOTHERAPY TX PLANNING SYSTEM | IWB | NUCLETRON CORP. | PLATO BPS | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | Other |