FDA Adverse Event Malfunction Summary report: N

PLATO BPS V. 13.2

MDR report key: 89529 · Received May 6, 1997

Report

Report Number
1121753-1997-00015
Event Type
Malfunction
Date Received
May 6, 1997
Report Date
April 17, 1997
Manufacturer
NUCLETRON CORP.
Product Code
IWB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
UNKNOWN

Narratives

Description of Event or Problem · 1

FOR SOFTWARE V. 13.2, IF AN EXISTING PLAN IS SELECTED, THEN READ IN A AS A CUSTOM FILE WITH DIFFERENT STEP SIZE, THE PROGRAM ALLOWS ONE TO WRITE A PROGRAM WITH A DIFFERENT STEP SIZE THAN THE STEP SIZE ORIGINALLY PROGRAMMED. THE JOURNAL FILE PRINTOUT DISPLAYS THE MESSAGES: (1) "INACCURATE DOSE CALCULATION MAY RESULT" AND (2) PROGRAM CARD IS DISABLED". HOWEVER THE PROGRAM CARD IS NOT DISABLED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PLATO BPS V. 13.2 RADIOTHERAPY TX PLANNING SYSTEM IWB NUCLETRON CORP. PLATO BPS *

Patients

Seq Age Sex Outcome Treatment
1 * Other