FDA Adverse Event
Malfunction
Summary report: N
NPS
MDR report key: 89485
·
Received May 6, 1997
Report
- Report Number
- 1121753-1997-00005
- Event Type
- Malfunction
- Date Received
- May 6, 1997
- Date of Event
- September 17, 1996
- Report Date
- April 9, 1997
- Manufacturer
- NUCLETRON CORP.
- Product Code
- IWB
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OK, US
- Reporter Occupation
- PHYSICIAN
Narratives
Description of Event or Problem · 1
FOR SOFTWARE SEED VERSIONS V. 11.0 AND V. 11.1, A SOURCE DECAY CALCULATION IS NOT PERFORMED WHEN AN EXISTING PATIENT FILE IS RECALLED AND WHEN THE PATIENT TREATMENT PARAMETERS CONTAINED IN THAT FILE ARE UNCHANGED. IN THIS SITUATION, THE ORIGINAL TREATMENT PARAMETERS ARE DISPLAYED, INCLUDING THE ORIGINAL SOURCE STRENGTH, AND A SOURCE DECAY CALCULATION TO CORRECT SOURCE DAY TO THE ACTUAL TIME OF TREATMENT WILL NOT BE UTILIZED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | NPS | RADIOTHERAPY TREATMENT PLANNING SYSTEM | IWB | NUCLETRON CORP. | MLDR/SEED V 11.1 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * | Other |