FDA Adverse Event Malfunction Summary report: N

NPS

MDR report key: 89485 · Received May 6, 1997

Report

Report Number
1121753-1997-00005
Event Type
Malfunction
Date Received
May 6, 1997
Date of Event
September 17, 1996
Report Date
April 9, 1997
Manufacturer
NUCLETRON CORP.
Product Code
IWB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

FOR SOFTWARE SEED VERSIONS V. 11.0 AND V. 11.1, A SOURCE DECAY CALCULATION IS NOT PERFORMED WHEN AN EXISTING PATIENT FILE IS RECALLED AND WHEN THE PATIENT TREATMENT PARAMETERS CONTAINED IN THAT FILE ARE UNCHANGED. IN THIS SITUATION, THE ORIGINAL TREATMENT PARAMETERS ARE DISPLAYED, INCLUDING THE ORIGINAL SOURCE STRENGTH, AND A SOURCE DECAY CALCULATION TO CORRECT SOURCE DAY TO THE ACTUAL TIME OF TREATMENT WILL NOT BE UTILIZED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NPS RADIOTHERAPY TREATMENT PLANNING SYSTEM IWB NUCLETRON CORP. MLDR/SEED V 11.1 *

Patients

Seq Age Sex Outcome Treatment
1 * Other