FDA Adverse Event Malfunction Summary report: N

ECHOTIP ULTRA ENDOSCOPIC ULTRASOUND NEEDLE

MDR report key: 8944399 · Received August 29, 2019

Report

Report Number
3001845648-2019-00442
Event Type
Malfunction
Date Received
August 29, 2019
Date of Event
April 25, 2019
Report Date
May 31, 2019
Manufacturer
COOK IRELAND LTD
Product Code
FCG
UDI-DI
00827002315204
PMA / PMN Number
K083330
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

PMA/510(K) # K080330. 1 UNIT OF LOT NUMBER C1534336 OF ECHO-19 WAS RETURNED OPENED IN ITS ORIGINAL PACKAGING. THE DEVICE RELATED TO THIS OCCURRENCE UNDERWENT A LABORATORY EVALUATION. THE TIP OF THE NEEDLE WAS FOUND TO BE DEFORMED/BENT. PRIOR TO DISTRIBUTION, ALL ECHO-19 DEVICES ARE SUBJECTED TO FUNCTIONAL CHECKS AND VISUAL INSPECTION TO ENSURE DEVICE INTEGRITY. THESE INSPECTIONS AND FUNCTIONAL CHECKS ARE OUTLINED IN INTERNAL PROCEDURES IN PLACE AT CIRL. A REVIEW OF THE MANUFACTURING RECORDS FOR ECHO-19 OF LOT NUMBER C1534336 DID NOT REVEAL ANY DISCREPANCIES THAT COULD HAVE CONTRIBUTED TO THIS COMPLAINT ISSUE. THE REVIEW OF RELEVANT MANUFACTURING RECORDS, CONFIRMS THE FAILURE MODE HAS NOT PREVIOUSLY OCCURRED WITH THE CURRENT LOT NUMBER. BASED ON THE INFORMATION AVAILABLE TO DATE, THERE IS NO EVIDENCE TO SUGGEST THAT THERE ARE ANY MANUFACTURING ISSUES ASSOCIATED WITH LOT NUMBER C1534336. THE NOTES SECTION OF THE INSTRUCTIONS FOR USE, WHICH ACCOMPANIES THIS DEVICE INSTRUCTS THE USER TO; "VISUALLY INSPECT WITH PARTICULAR ATTENTION TO KINKS, BENDS AND BREAKS. IF AN ABNORMALITY IS DETECTED THAT WOULD PROHIBIT PROPER WORKING CONDITION, DO NOT USE". THERE IS NO EVIDENCE TO SUGGEST THAT THE CUSTOMER DID NOT FOLLOW THE INSTRUCTIONS FOR USE. THE FAILURE OF NEEDLE TIP BENT WAS OBSERVED IN THE LABORATORY. A DEFINITIVE ROOT CAUSE FOR THE CUSTOMER COMPLAINT COULD NOT BE DETERMINED AS CIRCUMSTANCES OF USE CANNOT BE REPLICATED IN THE LABORATORY. A POSSIBLE ROOT CAUSE COULD BE ATTRIBUTED TO ADVANCEMENT INTO A HARD LESION OR PRESENCE OF BIOLOGICAL MATTER COULD CONTRIBUTE TO THE ADVANCEMENT DIFFICULTY. COMPLAINT IS CONFIRMED AS THE FAILURE WAS VERIFIED IN THE LABORATORY. ACCORDING TO THE INITIAL REPORTER, THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE. COMPLAINTS OF THIS NATURE WILL CONTINUE TO BE MONITORED FOR POTENTIAL EMERGING TRENDS.

Description of Event or Problem · 1

NEEDLE CANNOT BE ADVANCED IN ENDOSCOPY. USER TRIED SEVERAL TIMES BUT FAILED. USER CHANGED ANOTHER SAME DEVICE TO COMPLETE THE BIOPSY. DEVICE EVALUATED AND A DISTAL KINK ON THE NEEDLE WAS OBSERVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
737914 ECHOTIP ULTRA ENDOSCOPIC ULTRASOUND NEEDLE FCG KIT, NEEDLE, BIOPSY FCG COOK IRELAND LTD G31520 C1534336 00827002315204

Patients

Seq Age Sex Outcome Treatment
1 62 YR