FDA Adverse Event Malfunction Summary report: N

I-STAT TROPONIN (CTNI) CARTRIDGE

MDR report key: 8939153 · Received August 28, 2019

Report

Report Number
2245578-2019-00212
Event Type
Malfunction
Date Received
August 28, 2019
Date of Event
June 10, 2019
Report Date
September 11, 2019
Manufacturer
ABBOTT POINT OF CARE
Product Code
MMI
UDI-DI
10054749001986
PMA / PMN Number
K031739
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

APOC INCIDENT #: (B)(4). THE INVESTIGATION WAS COMPLETED ON 09/10/2019. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THAT THE CARTRIDGE LOT MET FINISHED GOODS (FG) RELEASE CRITERIA. RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE OUTLINED IN APPENDIX 1 OF Q04.01.003 REV. AD (PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE). NO DEFICIENCY HAS BEEN IDENTIFIED FOR CTNI CARTRIDGE LOT K19043A.

Additional Manufacturer Narrative · 1

(B)(4). APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.

Description of Event or Problem · 1

ON 08/15/2019, ABBOTT POINT OF CARE (APOC) WAS CONTACTED BY A CUSTOMER REGARDING I-STAT TROPONIN CARTRIDGES THAT YIELDED A SUSPECTED DISCREPANT TROPONIN RESULTS ON A (B)(6) MALE PATIENT. THERE WAS NO ADDITIONAL PATIENT INFORMATION AT THE TIME OF THIS REPORT. RETURN PRODUCT IS NOT AVAILABLE FOR INVESTIGATION. METHOD: DATE; COLLECTED; TESTED: RESULT: POSITIVE CUT-OFF: VALUE: I-STAT, (B)(6) 2019, 09:51, 09:51, 0.01 NG/ML, 0.08; DIMENSION EXL, (B)(6) 2019, 09:40, 09:45, 0.196 NG/ML, 0.056. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON THE LIMITED INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY. PER I-STAT SYSTEM MANUAL: ART: 714258-00Q REPORTABLE RANGE: 0.00 - 50.00 REFERENCE RANGE: 0.00 - 0.03 (REPRESENTS THE 0 TO 97.5% RANGE OF RESULTS), REFERENCE RANGE: 0.00 - 0.08 (REPRESENTS THE 0 TO 99% RANGE OF RESULTS).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
736830 I-STAT TROPONIN (CTNI) CARTRIDGE CTNI CARTRIDGE MMI ABBOTT POINT OF CARE NA K19043A 10054749001986

Patients

Seq Age Sex Outcome Treatment
1 60 YR