FDA Adverse Event Malfunction Summary report: N

EVIS EXERA LLL COLONOVIDEOSCOPE

MDR report key: 8938439 · Received August 28, 2019

Report

Report Number
8010047-2019-03101
Event Type
Malfunction
Date Received
August 28, 2019
Date of Event
August 2, 2019
Report Date
September 18, 2019
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORP.
Product Code
FDF
UDI-DI
04953170305177
PMA / PMN Number
K131780
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE SUBJECT DEVICE HAS NOT BEEN RETURNED TO OMSC BUT WAS RETURNED TO OLYMPUS (B)(4). (B)(4) SENT THE SUBJECT DEVICE TO A THIRD PARTY LABORATORY FOR MICROBIOLOGICAL TESTING. AS A RESULT OF THE TESTING, THE SAMPLE COLLECTED FROM ALL CHANNELS OF THE DEVICE TESTED POSITIVE FOR MICROCOCCACEAE (1CFU / ENDOSCOPE). THE TESTING RESULT CLEARED THE FRENCH GUIDELINE. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE SUPPLEMENTED.

Additional Manufacturer Narrative · 1

THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO PROVIDE ADDITIONAL INFORMATION. OMSC REVIEWED THE MANUFACTURING HISTORY (DHR) OF THE SUBJECT DEVICE AND CONFIRMED NO IRREGULARITY. THE EXACT CAUSE OF THE REPORTED EVENT COULD NOT BE CONCLUSIVELY DETERMINED AT THIS TIME. IF ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE SUPPLEMENTED.

Description of Event or Problem · 1

OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT AS A RESULT OF MICROBIOLOGICAL TESTING BY THE USER FACILITY, FOLLOWING MICROBES WERE DETECTED FROM THE SAMPLE COLLECTED FROM THE SUBJECT DEVICE. -THE INSTRUMENT CHANNEL: PSEUDOMONAS AERUGINOSA AND STENOTROPHOMONAS MALTOPHILIA (THE TOTAL CFU OF THE PSEUDOMONAS AERUGINOSA AND STENOTROPHOMONAS MALTOPHILIA IS 45CFU) -THE AUXILIARY CHANNEL: PSEUDOMONAS AERUGINOSA AND STENOTROPHOMONAS MALTOPHILIA (THE TOTAL CFU OF THE PSEUDOMONAS AERUGINOSA AND STENOTROPHOMONAS MALTOPHILIA IS 33CFU) -THE AIR/WATER CHANNEL: PSEUDOMONAS AERUGINOSA, STENOTROPHOMONAS MALTOPHILIA, AND BACILLUS CEREUS (THE TOTAL CFU OF THE PSEUDOMONAS AERUGINOSA, STENOTROPHOMONAS MALTOPHILIA, AND BACILLUS CEREUS IS 9CFU) -THE SUCTION CHANNEL: PSEUDOMONAS AERUGINOSA AND STENOTROPHOMONAS MALTOPHILIA (THE TOTAL CFU OF THE PSEUDOMONAS AERUGINOSA AND STENOTROPHOMONAS MALTOPHILIA IS >150CFU) THE DEVICE HAD BEEN REPROCESSED WITH A NON-OLYMPUS AUTOMATED ENDOSCOPE REPROCESSOR, SOLUSCOPE S4, USING PERACETIC ACID. THERE WAS NO REPORT OF INFECTION ASSOCIATED WITH THIS REPORT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
732440 EVIS EXERA LLL COLONOVIDEOSCOPE COLONOVIDEOSCOPE FDF OLYMPUS MEDICAL SYSTEMS CORP. CF-H190I 04953170305177

Patients

Seq Age Sex Outcome Treatment
1