FDA Adverse Event Malfunction Summary report: N

SKYTRON

MDR report key: 893497 · Received August 7, 2007

Report

Report Number
1825014-2007-00008
Event Type
Malfunction
Date Received
August 7, 2007
Date of Event
July 19, 2007
Report Date
July 25, 2007
Manufacturer
THE KMW GROUP, SKYTRON DIVISION
Product Code
BWN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TABLE WAS EXAMINED IN 2007. UPON INSPECTION, IT WAS IMMEDIATELY NOTICEABLE THAT THERE WAS DAMAGE TO THE LOW VOLTAGE (24 VOLT) CIRCUIT CONDUCTORS. THESE CONDUCTORS CAN BE FOUND NEAR THE ELECTRICAL CONNECTOR THAT PLUGS INTO THE TABLES RELAY BOX ASSEMBLY. THE EXTERIOR INSULATION OF THE CONDUCTORS EXHIBITED SIGNS OF DAMAGE FROM HEAT (SLIGHTLY MELTED). ADDITIONAL INVESTIGATION REVEALED THAT THE CAUSE OF THE FAILURE WAS A SHORT TO GROUND. IT IS APPARENT ON THE RED (+24 VOLT) CONDUCTOR THAT IT WAS PINCHED AGAINST A METAL BRACKET CREATING A RESISTANT POINT. THIS IS EVIDENT BECAUSE THE IMPRINT OF THE METAL BRACKET IS SHAPED INTO THE OUTER JACKET OF THE CONDUCTOR. DAMAGE TO THE RELAY BOX CONTROL CIRCUIT RESULTED FROM THE SHORT WHICH MADE THE TABLE INOPERABLE. IT APPEARS THAT THE LAST INDIVIDUAL WHO SERVICED THE TABLE OR REMOVED THE ACCESS COVER DID NOT EXERCISE CARE TO ROUTE THE WIRE HARNESS CORRECTLY AND CREATED A PINCH POINT.

Description of Event or Problem · 1

PT WAS ON A SKYTRON 3600B TABLE WHEN IT STARTED TO SMELL LIKE SMOKE. THE PT WAS ANESTHETIZED AND WAS MOVED TO ANOTHER TABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SKYTRON SURGICAL TABLE BWN THE KMW GROUP, SKYTRON DIVISION 3600B

Patients

Seq Age Sex Outcome Treatment
1 YR