FDA Adverse Event Malfunction Summary report: N

AURICAL AUD/PMM, 1081

MDR report key: 8934276 · Received August 27, 2019

Report

Report Number
9612197-2019-00008
Event Type
Malfunction
Date Received
August 27, 2019
Date of Event
August 6, 2019
Report Date
September 23, 2019
Manufacturer
NATUS MEDICAL DENMARK (GN OTOMETRICS)
Product Code
EWO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 0

PRODUCT WAS RETURNED AND EVALUATED. PROBLEM DESCRIPTION: INTERMITTENT CONNECTION - FAILED DURING TIME TEST. OVER HEATING - UNABLE TO REPRODUCE ERROR. FREEFIT BATTERY COVER DAMAGED. RESOLUTION: SERVICE REPLACEMENT UNIT. TRANSFERRED CALIBRATION SETTINGS. REPLACED FREEFIT BATTERY COVER. DEVICE REPAIRED AND SHIPPED BACK TO THE CUSTOMER.

Additional Manufacturer Narrative · 1

ISSUE AFFECTING AURICAL AUD/PMM, 1081, SN: (B)(4). DEVICE DISCONNECTED & BURNING SMELL. PATIENT INFORMATION - NO PATIENT INVOLVEMENT. PMA/510(K) - THIS DEVICE IS EXEMPT.

Description of Event or Problem · 1

ISSUE AFFECTING AURICAL AUD/PMM, 1081, SN: (B)(4). DEVICE DISCONNECTED & BURNING SMELL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
728006 AURICAL AUD/PMM, 1081 AURICAL AUD/PMM, 1081 EWO NATUS MEDICAL DENMARK (GN OTOMETRICS) 8-04-14500

Patients

Seq Age Sex Outcome Treatment
1 Other