INCLUSIVE TAPERED IMPLANT 4.7MMD X 11.5MML X 4.5MMP

MDR report key: 8933284 · Received August 26, 2019

Report

Report Number
3011649314-2019-00416
Event Type
Injury
Date Received
August 26, 2019
Date of Event
May 1, 2019
Report Date
November 15, 2019
Manufacturer
PRISMATIK DENTALCRAFT, INC.
Product Code
DZE
PMA / PMN Number
K121406
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TU
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

CORRECTION: DEVICE BRAND CORRECTED TO INCLUSIVE TAPERED IMPLANT 4.7MMD X 11.5MML X 4.5MMP. THE DEVICE INVESTIGATION HAS BEEN COMPLETED AND THE RESULTS ARE AS FOLLOWS: DEVICE HISTORY RECORD THE DHR WAS REVIEWED AND THERE WAS NO EVIDENCE DISCOVERED TO INDICATE THAT A PRODUCT DEFECT OR NON-CONFORMITY CONTRIBUTED TO THE ISSUE. THE PART MET ALL THE CRITERIA CALLED FOR IN THE PRODUCTION ROUTER. RETURNED SAMPLE: CUSTOMER DID NOT RETURN THE COMPLAINT PART FOR INVESTIGATION. ROOT CAUSE: "LOST OSSEOINTEGRATION" IS A COMMON COMPLAINT IN REGARDS TO IMPLANT FAILURE. THIS OCCURS WHEN THE PATIENT'S BONE DOES NOT INTEGRATE WITH THE IMPLANT SURFACE. THE POSSIBLE RESPONSES TO THIS COMPLAINT COULD BE ATTRIBUTED TO VARIOUS CAUSES MENTIONED BELOW. ALTHOUGH THE ROOT CAUSE FOR LOST OSSEOINTEGRATION IS INCONCLUSIVE AND SPECIFIC TO EACH CASE, PROBABLE CAUSES COULD BE THE LOSS OF PRIMARY STABILITY AT THE OSTEOTOMY SITE DUE TO INSUFFICIENT BONE OR POOR BONE QUALITY; EITHER THE BONE WAS TOO SOFT OR THE OPERATOR ERRED IN CREATING AN OSTEOTOMY BIGGER THAN THE SIZE OF THE IMPLANT DIAMETER. PREMATURE LOADING, PATIENT'S HEALTH, PERI-IMPLANTITIS, SMOKING, AND LACK OF ORAL HYGIENE MAY ALSO BE CONTRIBUTING FACTORS TO THE LACK OF OSSEOINTEGRATION. IFU 3023579 REV 3.0 (INCLUSIVE DENTAL IMPLANT SYSTEM) CONTAINS THE FOLLOWING STATEMENT IN WARNING SECTION: "ABSOLUTE SUCCESS CANNOT BE GUARANTEED. FACTORS SUCH AS INFECTION, DISEASE AND INADEQUATE BONE QUALITY AND/OR QUANTITY CAN RESULT IN OSSEOINTEGRATION FAILURES FOLLOWING SURGERY OR INITIAL OSSEOINTEGRATION." THE IFU ALSO CONTAINS THE FOLLOWING STATEMENT IN PRECAUTION SECTION: "MINIMIZING TISSUE DAMAGE IS CRUCIAL TO SUCCESSFUL IMPLANT OSSEOINTEGRATION. IN PARTICULAR, CARE SHOULD BE TAKEN TO ELIMINATE SOURCES OF INFECTION, CONTAMINANTS, SURGICAL AND THERMAL TRAUMA. RISK OF OSSEOINTEGRATION FAILURE INCREASES AS TISSUE TRAUMA INCREASES. ALL DRILLING PROCEDURES SHOULD BE PERFORMED AT 2000 RPM OR LESS UNDER CONTINUAL AND COPIOUS IRRIGATION. ALL SURGICAL INSTRUMENTS USED MUST BE IN GOOD CONDITION AND SHOULD BE USED CAREFULLY TO AVOID DAMAGE TO IMPLANTS OR OTHER COMPONENTS. IMPLANTS SHOULD BE PLACED WITH SUFFICIENT STABILITY; HOWEVER, EXCESSIVE INSERTION TORQUE MAY RESULT IN IMPLANT FRACTURE, OR FRACTURE OR NECROSIS OF THE IMPLANT SITE. THE PROPER SURGICAL PROTOCOL SHOULD BE STRICTLY ADHERED TO." SMOKING IS KNOWN TO INHIBIT LOCAL WOUND HEALING AND MAY AFFECT SURVIVAL OF IMPLANTS. SMOKING HAS A STRONG INFLUENCE ON THE COMPLICATION RATES OF IMPLANTS. IT CAUSES SIGNIFICANTLY MORE MARGINAL BONE LOSS AFTER IMPLANT PLACEMENT, INCREASED THE INCIDENCE OF PERI-IMPLANTITIS (DEEP MUCOSAL POCKETS AROUND DENTAL IMPLANTS, INFLAMMATION OF THE PERI-IMPLANT MUCOSA, AND INCREASED RESORPTION OF PERI-IMPLANT BONE, AND AFFECTS THE SUCCESS RATES OF BONE GRAFTS.

Additional Manufacturer Narrative · 1

THE DEVICE HAS NOT BEEN RETURNED. IF/WHEN THE DEVICE IS RETURNED AN INVESTIGATION WILL BE CARRIED OUT AND A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE PATIENT'S RACE AND ETHNICITY WERE NOT PROVIDED; HOWEVER, THE PATIENT'S NATIONALITY IS LISTED AS (B)(6).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INCLUSIVE TAPERED IMPLANT FAILED. THE PATIENT HAS BONE GRADE TYPE II. THE PATIENT HAS A HISTORY OF SMOKING AND WEAK BONE. THE PATIENT PRESENTED ON (B)(6)2019 FOR PRIMARY PROCEDURE ON TOOTH #28 (UNIVERSAL). ON (B)(6) 2019 THE PATIENT PRESENTED FOR SECOND STAGE OF THE PROCEDURE WITH COMPLAINTS OF PAIN. UPON EXAMINATION, THE PROVIDER NOTES AN INFECTION AND A LOSS OF OSSEOINTEGRATION. IT WAS AT THAT TIME THE DEVICE WAS REMOVED. THE PROVIDER STATES THE PATIENT CURRENT STATUS, "FINE NOW".

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
723247 INCLUSIVE TAPERED IMPLANT 4.7MMD X 11.5MML X 4.5MMP INCLUSIVE TAPERED IMPLANT DZE PRISMATIK DENTALCRAFT, INC. 70-1070-IMP0012 6024271

Patients

Seq Age Sex Outcome Treatment
1 41 YR Required Intervention