FDA Adverse Event Malfunction Summary report: N

BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES

MDR report key: 8931139 · Received August 26, 2019

Report

Report Number
1024879-2019-01514
Event Type
Malfunction
Date Received
August 26, 2019
Date of Event
August 8, 2019
Report Date
October 8, 2019
Manufacturer
BECTON, DICKINSON & CO., (BD)
Product Code
JKA
UDI-DI
50382903679862
PMA / PMN Number
BK050036
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

INVESTIGATION SUMMARY: BD RECEIVED SAMPLES AND PHOTOS FROM THE CUSTOMER FACILITY FOR INVESTIGATION. THE PHOTOS WERE EVALUATED AND THE CUSTOMER¿S INDICATED FAILURE MODE FOR FOREIGN MATERIAL WITH THE INCIDENT LOT WAS OBSERVED. ADDITIONALLY, EVALUATION OF THE CUSTOMER SAMPLES WAS PERFORMED AND A SMALL PIECE OF CARDBOARD WAS OBSERVED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT.

Description of Event or Problem · 0

MATERIAL NO. 367986 BATCH NO. 9030859. IT WAS REPORTED THAT BEFORE USE OF THE BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES THE TUBE GEL LOOKS LIKE IT HAS MOLD GROWING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: TUBE GEL LOOKS LIKE HAS MOLD GROWING ONLY SEEN 1 TUBE WITH THIS ISSUE. CUSTOMER STATES THAT THERE WAS ONLY 1 TUBE IN THE REMAINING CASE OF SST TUBES WITH MOLD IN THE GEL. THE OTHER TUBES APPEAR NORMAL, AND THEY ARE USING THEM WITH NO PROBLEMS.

Additional Manufacturer Narrative · 1

A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION AND/OR DEVICE HISTORY REVIEW, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

MATERIAL NO. 367986, BATCH NO. 9030859. IT WAS REPORTED THAT BEFORE USE OF THE BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES THE TUBE GEL LOOKS LIKE IT HAS MOLD GROWING. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: TUBE GEL LOOKS LIKE HAS MOLD GROWING ONLY SEEN 1 TUBE WITH THIS ISSUE. CUSTOMER STATES THAT THERE WAS ONLY 1 TUBE IN THE REMAINING CASE OF SST TUBES WITH MOLD IN THE GEL. THE OTHER TUBES APPEAR NORMAL, AND THEY ARE USING THEM WITH NO PROBLEMS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
727086 BD VACUTAINER® SST¿ BLOOD COLLECTION TUBES BLOOD SPECIMEN COLLECTION DEVICE JKA BECTON, DICKINSON & CO., (BD) 9030859 50382903679862

Patients

Seq Age Sex Outcome Treatment
1 Other