INRATIO
Report
- Report Number
- 2954730-2007-00363
- Event Type
- Malfunction
- Date Received
- August 10, 2007
- Date of Event
- August 2, 2007
- Report Date
- August 6, 2007
- Manufacturer
- HEMOSENSE, INC.
- Product Code
- JPA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- UNKNOWN
Narratives
INRATIO PRECISION DATA PROVIDED BY END-USER : FIRST TEST INR = 1.5 SECOND TEST INR = 1.1 MEAN = 1.3 ; SD = 0.28; %CV = 22%. THE %CV IS GREATER THAN 20%. PER INTERNAL PROCEDURE, THE PRECISION FAILED THE CRITERIA FOR PRECISION. PRODUCTS WILL BE TESTED. PER INTERNAL PROTOCOL, IN-HOUSE RETAIN STRIPS 060656 WERE TESTED FOR PRECISION. THE ACCEPTANCE CRITERIA IS AS FOLLOWS: IF THE %CV IS LESS THAN OR EQUAL TO 16%, THEN IT MEETS THE CRITERIA FOR PRECISION. IF BOTH SAMPLES PASS FOR EACH LOT THEN NO FURTHER ACTION IS REQUIRED. IF ONE LOT FAILS, THEN ADDITIONAL TESTING WILL BE PERFORMED WITH 2 MORE NORMAL DONORS. IF BOTH SAMPLES FAIL ON ANY LOT OR IF ANY LOT FAILS ADDITIONAL TESTING, THEN A REVIEW OF TRENDING DATA WILL BE PERFORMED AND A COURSE OF ACTION TAKEN. RESULT OF RETAINED STRIPS TEST ( LOT 060656) PERFORMED IN 2007 AS FOLLOWS: LOT 060656 PT. #1: NORMAL 1.0, NORMAL 1.2, MEAN 1.2, SD 1.1, %CV 0.14%. PT #2: NORMAL 1.0, NORMAL 1.0, MEAN 1.0, SD 0.0, %CV 0.0%. BASED ON THE ABOVE TEST RESULTS, RETAINED STRIPS LOT 060656 MEETS THE CRITERIA FOR STRIP PRECISION.
CALLER ALLEGED IMPRECISION WITH INRATIO METER. RESULTS AS FOLLOWS: FIRST TEST INR = 1.5, SECOND TEST INR = 1.1.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | INRATIO | PROTHROMBIN TIME TEST | JPA | HEMOSENSE, INC. | 060656 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | YR |