WAYPOINT STEREOTACTIC PLATFORM
Report
- Report Number
- 3002250546-2019-00006
- Event Type
- Malfunction
- Date Received
- August 23, 2019
- Date of Event
- July 24, 2019
- Report Date
- August 23, 2019
- Manufacturer
- FHC, INC,
- Product Code
- HAW
- UDI-DI
- 00873263005903
- PMA / PMN Number
- K092192
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- ME, US
- Reporter Occupation
- 003
Narratives
DEVICE EVALUATION: THE PLATFORM USED IN THE SURGERY WAS RETURNED AND EVALUATED. THE INVESTIGATION SHOWS THAT NO ISSUES WERE FOUND DURING RECEIVING, PROCESSING, MANUFACTURING AND QC'ING OF THE ORIGINAL PLATFORM. THE RETURNED PLATFORM WAS INSPECTED AND FOUND TO HAVE NON-CONFORMING MEASUREMENTS THAT ARE WELL OVER THE SPECIFIED TOLERANCES. THIS IS INDICATIVE THAT THE GEOMETRY OF THE PLATFORM WAS ALTERED AFTER LEAVING FHC. THE FOLLOW-UP CASE WAS SUCCESSFUL. SINCE THE SECOND PLATFORM WAS PROCESSED, MANUFACTURED AND QC'D JUST AS THE ORIGINAL, THIS LENDS TO THE CONCLUSION THAT THE PROBLEM WITH THE FIRST PLATFORM HAPPENED OUTSIDE OF FHC CONTROL. A ROOT CAUSE COULD NOT BE DETERMINED AT THIS TIME.
PHYSICIAN HAD DIFFICULTY ATTACHING THE STARFIX PLATFORM TO THE IMPLANTED FIDUCIALS AT THE START OF THE CASE. HE CHECKED MULTIPLE TIMES THE PEEK STANDOFFS FROM THE SKS KITS AND WAS STILL HAVING DIFFICULTY ATTACHING THE FINAL POSTERIOR LEFT FOOT THE STANDOFF. IT SHOWED A 3MM POSTERIOR SHIFT THAT PHYSICIAN DID NOT FEEL COMFORTABLE PROCEEDING WITH THE SURGERY WITH 3 LEGS WITHOUT KNOWING WHY THE 4TH LEG WAS SHIFTED SO MUCH. THE SURGERY WAS CANCELLED AND THE IMPACT TO THE PATIENT WAS OPENING THE FIDUCIAL SITES AND CLOSING AGAIN AND HAVING THE PATIENT GET ANOTHER CT TO DOUBLE CHECK EVERYTHING. THE CASE CONCLUDED SUCCESSFULLY WITH THE USE OF THE PLATFORM CREATED FROM THE 2ND SET OF CT SCANS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 722120 | WAYPOINT STEREOTACTIC PLATFORM | WAYPOINT STEREOTACTIC SYSTEM | HAW | FHC, INC, | MP-KIT-P-BI | 00873263005903 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |