FDA Adverse Event Malfunction Summary report: N

STEALTHSTATION S8 PREMIUM SYSTEM

MDR report key: 8923835 · Received August 22, 2019

Report

Report Number
1723170-2019-04611
Event Type
Malfunction
Date Received
August 22, 2019
Date of Event
August 6, 2019
Report Date
October 3, 2019
Manufacturer
MEDTRONIC NAVIGATION, INC
Product Code
HAW
PMA / PMN Number
K162309
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 1

OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: 9735821, SERIAL/LOT #: UNKNOWN, UBD: UNKNOWN, UDI#: UNKNOWN. A MEDTRONIC REPRESENTATIVE WENT TO THE SITE TO TEST THE EQUIPMENT. TESTING REVEALED THAT PARTS WERE REPLACED. THE SYSTEM THEN PASSED THE SYSTEM CHECKOUT AND WAS FOUND TO BE FULLY FUNCTIONAL. NO PARTS HAVE BEEN RECEIVED BY THE MANUFACTURER FOR EVALUATION. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION: LOT NUMBER FOR PRODUCT ID: 9735821 IS P901711. THE PSU WAS RETURNED TO THE MANUFACTURER FOR ANALYSIS. ANALYSIS FOUND THAT THE REPORTED ISSUE COULD BE CONFIRMED. A CHECK OF THE LOGS FOUND A FIRMWARE INCOMPATIBILITY ISSUE AND A BUMP AND TEMPERATURE FAILURE. THE PSU ALSO FAILED AN ACCURACY TEST. ANALYSIS FOUND THAT THE REPORTED EVENT WAS RELATED TO AN ELECTRICAL ISSUE. THIS ISSUE WAS DOCUMENTED IN A MEDTRONIC NAVIGATION HARDWARE ANOMALY TRACKING DATABASE. IF INFORMATION IS PROVIDED IN THE FUTURE, A SUPPLEMENTAL REPORT WILL BE ISSUED.

Description of Event or Problem · 1

MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM. IT WAS REPORTED THAT THE POSITIONING SENSOR UNIT (PSU) COULD NOT BE RECOGNIZED BY THE BUMP OF PSU AND THE TEMPERATURE SENSOR DETECTION. THE ISSUE WAS RESOLVED AFTER REPLACING WITH A BACK-UP PRODUCT. THERE WAS NO PATIENT PRESENT WHEN THIS ISSUE WAS IDENTIFIED. AN UPDATE WAS PROVIDED THE ISSUE RESOLVED BY REPLACING THE PSU.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
714087 STEALTHSTATION S8 PREMIUM SYSTEM INSTRUMENT, STEREOTAXIC HAW MEDTRONIC NAVIGATION, INC 9735665

Patients

Seq Age Sex Outcome Treatment
1