FDA Adverse Event Injury Summary report: N

ALLEN MEDICAL SYSTEMS

MDR report key: 892305 · Received August 10, 2007

Report

Report Number
1221538-2007-00004
Event Type
Injury
Date Received
August 10, 2007
Date of Event
July 10, 2007
Report Date
August 10, 2007
Manufacturer
ALLEN MEDICAL SYSTEMS, INC.
Product Code
FWZ
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE BEACH CHAIR WAS NOT RETURNED FOR EVALUATION AS STATEMENTS FROM THE USER STAFF AND ALL VISUAL INDICATIONS WERE THAT A) THERE WAS NO DEVICE FAILURE, AND B) THAT THE DEVICE NEVER COLLIDED WITH THE FLOOR OR OTHERWISE SUSTAINED ANY DAMAGE. ALLEN MEDICAL CONCLUDED THE ROOT CAUSE OF THE INCIDENT WAS MISUSE BY THE HOSPITAL STAFF. TWO NURSES THERE AND THE OR DIRECTOR ADMITTED AT THAT TIME THAT THE TABLE WAS NOT SETUP PROPERLY FOR THE PATIENT PROCEDURE. FOLLOWING THE INCIDENT, THE ALLEN REP PROVIDED A CORRECTIVE IN-SERVICE TRAINING WITH THE STAFF. AFTERWARDS, THE OR DIRECTOR STATED SHE WAS COMFORTABLE WITH HER STAFF'S KNOWLEDGE OF THE DEVICE AND ITS PROPER USAGE NOW. THE OR DIRECTOR AGREED WITH THE CONCLUSION OF THIS INVESTIGATION - THAT THE INCIDENT WAS THE DIRECT RESULT OF USER-ERROR/MISUSE - DURING A PHONE CONVERSATION CONDUCTED BY AN ALLEN QUALITY REPRESENTATIVE. THIS INCIDENT OCCURRED DESPITE THE PROVISION OF ADEQUATE USER INSTRUCTIONS AND AT LEAST THREE IN-SERVICES PROVIDED BY TRAINED ALLEN REPRESENTATIVES. THE FIRST OF THESE WAS PROVIDED BY FORMER ALLEN SALES REP (DATE UNKNOWN) AND TWO SUBSEQUENT IN-SERVICE TRAINING SESSIONS WITH CURRENT ALLEN REP.

Description of Event or Problem · 1

THE FOLLOWING TOOK PLACE DURING A PATIENT PROCEDURE IN 2007. IT WAS REPORTED THAT THE STAFF AT THE USER FACILITY PREPARED FOR A PATIENT PROCEDURE BY SETTING THE ALLEN LIFT ASSIST BEACH CHAIR ON THE FOOT SECTION OF THE OR TABLE. THIS PORTION OF THE TABLE IS NOT INTENDED TO SUSTAIN THE WEIGHT OF THE DEVICE AND THE PATIENT. DURING THE BEGINNING OF THE PROCEDURE, THE TABLE SECTION COLLAPSED AT THE HINGE. THIS CAUSED THE SECTION OF THE TABLE (WITH THE ATTACHED CHAIR) TO STRIKE A DOCTOR'S FOOT - CAUSING INJURY. THERE WAS NO INJURY TO THE PATIENT. IT WAS INITIALLY REPORTED THAT THE DOCTOR MAY HAVE BROKEN A BONE IN HER FOOT, BUT ACCORDING TO THE OR DIRECTOR, MEDICAL EVALUATION SHOWED THERE TO BE NO FRACTURES. THE FOOT WAS JUST BRUISED AND VERY TENDER. THE DOCTOR WAS ABLE TO WALK AND WAS STILL WORKING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ALLEN MEDICAL SYSTEMS ALLEN LIFT ASSIST BEACH CHAIR FWZ ALLEN MEDICAL SYSTEMS, INC. A-91000-A5 NA

Patients

Seq Age Sex Outcome Treatment
1 YR Required Intervention