FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 8922070 · Received August 22, 2019

Report

Report Number
3008642652-2019-06641
Event Type
Malfunction
Date Received
August 22, 2019
Date of Event
July 22, 2019
Report Date
August 21, 2019
Manufacturer
ZOLL MANUFACTURING CORPORATION
Product Code
MVK
UDI-DI
00855778005029
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION OF BATTERY CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (WON'T POWER ON) HAS BEEN CONFIRMED. UPON INVESTIGATION THE BATTERY CHARGER/MODEM WAS UNABLE TO POWER ON. THE CAUSE FOR THE FAILURE WAS ISOLATED TO A DEFECTIVE POWER SUPPLY BRICK. THE ROOT CAUSE FOR THE DEFECTIVE POWER SUPPLY BRICK COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY CHARGER/MODEM.

Description of Event or Problem · 1

A US DISTRIBUTOR REPORTED THAT A BATTERY CHARGER WOULD NOT POWER ON.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
716128 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL MANUFACTURING CORPORATION WCD 4000 00855778005029

Patients

Seq Age Sex Outcome Treatment
1