FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 8922070
·
Received August 22, 2019
Report
- Report Number
- 3008642652-2019-06641
- Event Type
- Malfunction
- Date Received
- August 22, 2019
- Date of Event
- July 22, 2019
- Report Date
- August 21, 2019
- Manufacturer
- ZOLL MANUFACTURING CORPORATION
- Product Code
- MVK
- UDI-DI
- 00855778005029
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION OF BATTERY CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (WON'T POWER ON) HAS BEEN CONFIRMED. UPON INVESTIGATION THE BATTERY CHARGER/MODEM WAS UNABLE TO POWER ON. THE CAUSE FOR THE FAILURE WAS ISOLATED TO A DEFECTIVE POWER SUPPLY BRICK. THE ROOT CAUSE FOR THE DEFECTIVE POWER SUPPLY BRICK COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY CHARGER/MODEM.
Description of Event or Problem · 1
A US DISTRIBUTOR REPORTED THAT A BATTERY CHARGER WOULD NOT POWER ON.