FDA Adverse Event Malfunction Summary report: N

EMAX 2 PLUS MOTOR

MDR report key: 8921360 · Received August 22, 2019

Report

Report Number
1045834-2019-54348
Event Type
Malfunction
Date Received
August 22, 2019
Date of Event
August 9, 2019
Report Date
August 12, 2019
Manufacturer
DEPUY SYNTHES PRODUCTS LLC
Product Code
HBC
UDI-DI
00845384002055
PMA / PMN Number
K080802
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

DEPUY SYNTHES IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH DEPUY SYNTHES HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, DEPUY SYNTHES OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. THE ACTUAL DEVICE WAS RETURNED FOR EVALUATION. DURING EVALUATION OF THE DEVICE, IT WAS DETERMINED THAT THE REPORTED CONDITION WAS NOT CONFIRMED. IT WAS FURTHER DETERMINED THAT THE DEVICE FAILED LOCTITE AND CABLE ASSESSMENT (CABLE TWISTED). THE ASSIGNABLE ROOT CAUSE WAS DETERMINED TO BE TRACED TO COMPONENT FAILURE WHICH IS NORMAL WEAR. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

Additional Manufacturer Narrative · 1

(B)(4). AS OF THIS DATE, THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE REPORTED CONDITION CANNOT BE CONFIRMED AND/OR DUPLICATED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SENT ACCORDINGLY.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PRE-SURGERY, IT WAS OBSERVED THAT THE MOTOR DEVICE WAS LEAKING OIL. THIS EVENT DID NOT OCCUR DURING SURGERY. THERE WAS NO PATIENT INVOLVEMENT. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH WILL BE SUBMITTED ACCORDINGLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
716846 EMAX 2 PLUS MOTOR MOTOR, DRILL, ELECTRIC - HANDPIECE HBC DEPUY SYNTHES PRODUCTS LLC 00845384002055

Patients

Seq Age Sex Outcome Treatment