FDA Adverse Event Malfunction Summary report: N

BD ULTRASAFE B100L NG GREEN BULK AMG

MDR report key: 8917573 · Received August 21, 2019

Report

Report Number
3009081593-2019-00196
Event Type
Malfunction
Date Received
August 21, 2019
Date of Event
August 2, 2019
Report Date
September 23, 2019
Manufacturer
BECTON DICKINSON HUNGARY KFT (BD)
Product Code
FMF
PMA / PMN Number
K972878
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 0

H.6. INVESTIGATION: UNCONFIRMED, NO SAMPLE RECEIVED. THE CUSTOMER ISSUED A COMPLAINT FOR A NEEDLE GUARD DAMAGED/DEFECTIVE DETECTED BY END USER. NEITHER SAMPLE NOR PHOTO WAS PROVIDED TO BD MEDICAL ¿ PHARMACEUTICAL SYSTEM (BDM-PS) FOR ANALYSIS. BDM-PS PERFORMED A BATCH HISTORY RECORD¿S REVIEW (BHR) INCLUDING A REVIEW OF ALL DATA COLLECTED DURING IN PROCESS AND QUALITY INSPECTIONS. THE BATCHES INVOLVED IN THIS COMPLAINT MEET ALL ACCEPTABLE QUALITY LEVELS (AQL¿S), WERE MANUFACTURED AND RELEASED ACCORDING TO APPLICABLE PROCEDURES AND SPECIFICATIONS. BASED ON THE INVESTIGATION CONCLUSION, BDM-PS WAS NOT ABLE TO CONFIRM THE SYMPTOM PERCEIVED BY THE CUSTOMER OR CORRELATE THIS SYMPTOM WITH A POTENTIAL CAUSE LINKED TO BD PROCESS.

Description of Event or Problem · 0

IT WAS REPORTED THAT BD ULTRASAFE B100L NG GREEN BULK AMG THE NEEDLE GUARD "MAY HAVE BEEN A LITTLE CROOKED" THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE GREEN NEEDLE GUARD MAY HAVE BEEN A LITTLE CROOKED.¿

Additional Manufacturer Narrative · 1

DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 8180960; MEDICAL DEVICE EXPIRATION DATE: 2022-06-30; DEVICE MANUFACTURE DATE: 2018-06-29. MEDICAL DEVICE LOT #: 8165754; MEDICAL DEVICE EXPIRATION DATE: 2022-05-31; DEVICE MANUFACTURE DATE: 2018-06-14. MEDICAL DEVICE LOT #: 8100792; MEDICAL DEVICE EXPIRATION DATE: 2022-03-31; DEVICE MANUFACTURE DATE: 2018-04-10. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION AND/OR DEVICE HISTORY REVIEW, A SUPPLEMENTAL REPORT WILL BE FILED.

Description of Event or Problem · 1

IT WAS REPORTED THAT BD ULTRASAFE B100L NG GREEN BULK AMG THE NEEDLE GUARD "MAY HAVE BEEN A LITTLE CROOKED" THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: THE GREEN NEEDLE GUARD MAY HAVE BEEN A LITTLE CROOKED.¿

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
712540 BD ULTRASAFE B100L NG GREEN BULK AMG PISTON SYRINGE FMF BECTON DICKINSON HUNGARY KFT (BD) SEE H.10

Patients

Seq Age Sex Outcome Treatment
1 Other