METRIX AIR
Report
- Report Number
- 1000113657-2019-00840
- Event Type
- Malfunction
- Date Received
- August 21, 2019
- Date of Event
- July 29, 2019
- Report Date
- May 14, 2020
- Manufacturer
- TRIVIDIA HEALTH, INC.
- Product Code
- NBW
- PMA / PMN Number
- K150052
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ
- Reporter Occupation
- OTHER
Narratives
(MANUFACTURER NARRATIVE = T, CORRECTED DATA = F) SECTIONS WITH ADDITIONAL INFORMATION AS OF 14-MAY-2020: H6: UPDATED FDA'S METHOD, RESULT, AND CONCLUSION CODES H10: METER WAS NOT RETURNED FOR EVALUATION. TEST STRIPS WERE NOT RETURNED FOR EVALUATION. RETENTION TESTING WAS PERFORMED USING TEST STRIPS FROM THE SAME LOT. RETENTION STRIP LOT TESTED AND PASSED. INTERNAL REPORT # (B)(4). METER AND TEST STRIPS WERE NOT RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-57: USER HAD AN INACCURATE REFERENCE: ALTERNATE METER: THE END USER IS COMPARING RESULTS OBTAINED FROM ONE OF TRIVIDIA'S BGM SYSTEM TO THE RESULTS FROM ANOTHER TRIVIDIA'S BGM SYSTEM. TEST STRIPS UDI: (B)(4). NOTE: MANUFACTURER CONTACTED CUSTOMER IN A FOLLOW-UP CALL TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - ABLE TO ESTABLISH CONTACT WITH CUSTOMER AND STATED REPLACEMENT PRODUCT RESOLVED INITIAL CONCERN.
CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS OBTAINED OF 248, 133, 152, 125 AND 118 MG/DL. CUSTOMER WAS ALSO CONCERNED WITH TEST RESULTS FROM METER TO METER COMPARISON OF 248 MG/DL USING PERSONAL METER AND 93 MG/DL USING ALTERNATIVE METER. CUSTOMER WAS USING EXPIRED TEST STRIPS WHEN TESTING WITH ALTERNATIVE METER (MANUFACTURER'S EXPIRATION DATE OF 12/22/2018). THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 98 - 113 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BACK TO BACK BLOOD TEST WAS PERFORMED BY THE CUSTOMER NON-FASTING AND PRODUCED TEST RESULTS OF 92 MG/DL AND 101 MG/DL USING PERSONAL METER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 11/26/2020 AND TEST STRIPS WERE OPENED ONE WEEK AGO. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: RESULT 1: 248MG/DL, DATE:(B)(6) 2019, TIME:12:01PM FASTING. RESULT 2: 133MG/DL, DATE:(B)(6) 2019, TIME:09:25PM FASTING. RESULT 3: 152MG/DL, DATE:(B)(6) 2019, TIME:12:00PM FASTING. RESULT 4: 125MG/DL, DATE:(B)(6) 2019, TIME:06:00PM FASTING. RESULT 5: 118MG/DL, DATE:(B)(6) 2019, TIME:07:29PM FASTING.
INTERNAL REPORT # (B)(4). METER AND TEST STRIPS WERE NOT RETURNED FOR EVALUATION. MOST LIKELY UNDERLYING ROOT CAUSE: MLC-57: USER HAD AN INACCURATE REFERENCE: ALTERNATE METER: THE END USER IS COMPARING RESULTS OBTAINED FROM ONE OF TRIVIDIA'S BGM SYSTEM TO THE RESULTS FROM ANOTHER TRIVIDIA'S BGM SYSTEM. TEST STRIPS UDI: (B)(4). NOTE: MANUFACTURER CONTACTED CUSTOMER IN A FOLLOW-UP CALL TO ENSURE THE REPLACEMENT PRODUCTS RESOLVED THE INITIAL CONCERN - ABLE TO ESTABLISH CONTACT WITH CUSTOMER AND STATED REPLACEMENT PRODUCT RESOLVED INITIAL CONCERN.
CONSUMER REPORTED COMPLAINT FOR HIGH BLOOD GLUCOSE TEST RESULTS. THE CUSTOMER IS CONCERNED WITH TEST RESULTS OBTAINED OF 248, 133, 152, 125 AND 118 MG/DL. CUSTOMER WAS ALSO CONCERNED WITH TEST RESULTS FROM METER TO METER COMPARISON OF 248 MG/DL USING PERSONAL METER AND 93 MG/DL USING ALTERNATIVE METER. CUSTOMER WAS USING EXPIRED TEST STRIPS WHEN TESTING WITH ALTERNATIVE METER (MANUFACTURER'S EXPIRATION DATE OF 12/22/2018). THE CUSTOMER'S EXPECTED FASTING BLOOD GLUCOSE TEST RESULT RANGE IS 98 - 113 MG/DL. THE CUSTOMER FEELS WELL AND DID NOT REPORT ANY SYMPTOMS. MEDICAL ATTENTION IS NOT REPORTED AS A RESULT OF THE ACTUAL BLOOD GLUCOSE RESULTS. DURING THE CALL, A BACK TO BACK BLOOD TEST WAS PERFORMED BY THE CUSTOMER NON-FASTING AND PRODUCED TEST RESULTS OF 92 MG/DL AND 101 MG/DL USING PERSONAL METER. THE PRODUCT IS STORED ACCORDING TO SPECIFICATION IN THE BEDROOM. THE TEST STRIP LOT MANUFACTURER'S EXPIRATION DATE IS 11/26/2020 AND TEST STRIPS WERE OPENED ONE WEEK AGO. THE METER MEMORY WAS REVIEWED FOR PREVIOUS TEST RESULT HISTORY: (B)(6).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 710130 | METRIX AIR | BLOOD GLUCOSE SYSTEM | NBW | TRIVIDIA HEALTH, INC. | METRIX AIR | MW3330S |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0 YR | SECOND THERAPY| SECOND THERAPY |