FDA Adverse Event
Malfunction
Summary report: N
FOCUS RADIATION TREATMENT PLANNING SYS
MDR report key: 89166
·
Received May 2, 1997
Report
- Report Number
- 1937649-1997-00003
- Event Type
- Malfunction
- Date Received
- May 2, 1997
- Date of Event
- April 18, 1997
- Report Date
- April 29, 1997
- Manufacturer
- COMPUTERIZED MEDICAL SYSTEMS, INC.
- Product Code
- KPQ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
CUSTOMER CREATED A TREATMENT PLAN DEFAULT BEAM SET-UP AND SAVED IT WITH INHOMOGENEITY CORRECTION TURNED "ON". SUBSEQUENTLY THE USER RECALLED THE STORED BEAM SET-UP INTO A PLAN WITH INHOMOGENEITY CORRECTION TURNED "OFF". THE TREATMENT PLAN CALCULATED WITH INHOMOGENEITY CORRECTION "ON". THIS RESULTED IN AN EFFECTIVE DEPTH GREATER THAN PHYSICAL DEPTH BECAUSE OF A HIGH DENSITY REGION IN THE PATH OF ONE BEAM. THE PLANNER NOTED THE DISCREPANCY AND NO PATIENT WAS TREATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FOCUS RADIATION TREATMENT PLANNING SYS | RADIATION TREATMENT PLANNING SYSTEM | KPQ | COMPUTERIZED MEDICAL SYSTEMS, INC. | 9200 | RELEASE 1.4.1 |