BD VACUTAINER® SST¿ BLOOD COLLECTION TUBE
Report
- Report Number
- 1024879-2019-01490
- Event Type
- Malfunction
- Date Received
- August 21, 2019
- Date of Event
- August 5, 2019
- Report Date
- October 8, 2019
- Manufacturer
- BECTON, DICKINSON & CO., (BD)
- Product Code
- JKA
- UDI-DI
- 50382903679862
- PMA / PMN Number
- BK050036
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION SUMMARY: BD RECEIVED SAMPLES FROM THE CUSTOMER FACILITY FOR INVESTIGATION. THE SAMPLES WERE TESTED/EVALUATED AND THE CUSTOMER'S INDICATED FAILURE MODE FOR STOPPER POP OFF WITH THE INCIDENT LOT WAS NOT OBSERVED AS ALL PRODUCT SPECIFICATIONS WERE MET. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT.
IT HAS BEEN REPORTED THAT ONE BD VACUTAINER® SST¿ BLOOD COLLECTION TUBE HAS BEEN FOUND WITH THE STOPPER POPPING OUT DURING USE. THE FOLLOWING HAS BEEN PROVIDED BY THE INITIAL REPORTER: -IT WAS REPORTED STOPPER POPPED OFF. EMAIL VERBIAGE RECEIVED, - "SST (GOLD) TUBE FOR CHEMISTRY TESTING- THE TOP COME OFF DURING DRAWING SAMPLE AND THEY HAVE TO TAKE ANOTHER TUBE TO COMPLETE SAMPLE DRAWING. THIS IS ALSO AN ONGOING ISSUE. SST (GOLD) ¿ LOT# 9042893, EXP: 1/31/2020"
DATE OF EVENT: UNKNOWN. THE DATE RECEIVED BY MANUFACTURER HAS BEEN USED FOR THIS FIELD. A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED, BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT HAS BEEN REPORTED THAT ONE BD VACUTAINER® SST¿ BLOOD COLLECTION TUBE HAS BEEN FOUND WITH THE STOPPER POPPING OUT DURING USE. THE FOLLOWING HAS BEEN PROVIDED BY THE INITIAL REPORTER: IT WAS REPORTED STOPPER POPPED OFF. EMAIL VERBIAGE RECEIVED, - "SST (GOLD) TUBE FOR CHEMISTRY TESTING- THE TOP COME OFF DURING DRAWING SAMPLE AND THEY HAVE TO TAKE ANOTHER TUBE TO COMPLETE SAMPLE DRAWING. THIS IS ALSO AN ONGOING ISSUE. SST (GOLD) ¿ LOT# 9042893, EXP: 1/31/2020."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 712670 | BD VACUTAINER® SST¿ BLOOD COLLECTION TUBE | BLOOD SPECIMEN COLLECTION DEVICE | JKA | BECTON, DICKINSON & CO., (BD) | 9042893 | 50382903679862 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |