FDA Adverse Event Injury Summary report: N

STYLE 15 SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 8915201 · Received August 21, 2019

Report

Report Number
9617229-2019-10260
Event Type
Injury
Date Received
August 21, 2019
Date of Event
July 19, 2019
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AZ
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: VISUAL ANALYSIS OF THE RETURNED DEVICE IDENTIFIED: ONE EXTENDED OPENING, CREASES AND YELLOW PARTICLES MATERIAL WAS FOUND ON GEL. A WEIGHT TEST OF THE DEVICE WAS VERIFIED AND THE DEVICE UNDERWEIGHT. A MICROSCOPIC ANALYSIS WAS PERFORMED WHICH IDENTIFY: ONE OPENING IN THE SHELL WITH SHARP EDGE WITH STRESS MARKS ASSESSED AS SURGICAL IMPACT OPENING. BASED ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS:A SHARP EDGE OPENING IN THE SHELL WITH STRESS MARKS DUE TO SURGICAL IMPACT OPENING.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED A LEFT SIDE RUPTURE. THE DEVICE HAS BEEN EXPLANTED.

Additional Manufacturer Narrative · 1

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCE'S NOTED. REASON FOR REOPERATION DUE TO RUPTURE. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. THIS IS A KNOWN POTENTIAL ADVERSE EVENT ADDRESSED IN THE PRODUCT LABELING.

Description of Event or Problem · 1

HEALTHCARE PROFESSIONAL REPORTED A LEFT SIDE RUPTURE. THE DEVICE HAS BEEN EXPLANTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
708731 STYLE 15 SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 2074755

Patients

Seq Age Sex Outcome Treatment
1 52 YR Required Intervention