STYLE 15 SILICONE GEL FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2019-10260
- Event Type
- Injury
- Date Received
- August 21, 2019
- Date of Event
- July 19, 2019
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FTR
- PMA / PMN Number
- P020056
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE EVALUATION: VISUAL ANALYSIS OF THE RETURNED DEVICE IDENTIFIED: ONE EXTENDED OPENING, CREASES AND YELLOW PARTICLES MATERIAL WAS FOUND ON GEL. A WEIGHT TEST OF THE DEVICE WAS VERIFIED AND THE DEVICE UNDERWEIGHT. A MICROSCOPIC ANALYSIS WAS PERFORMED WHICH IDENTIFY: ONE OPENING IN THE SHELL WITH SHARP EDGE WITH STRESS MARKS ASSESSED AS SURGICAL IMPACT OPENING. BASED ON THE DEVICE ANALYSIS THE FINAL ASSESSMENT IS:A SHARP EDGE OPENING IN THE SHELL WITH STRESS MARKS DUE TO SURGICAL IMPACT OPENING.
HEALTHCARE PROFESSIONAL REPORTED A LEFT SIDE RUPTURE. THE DEVICE HAS BEEN EXPLANTED.
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCE'S NOTED. REASON FOR REOPERATION DUE TO RUPTURE. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. THIS IS A KNOWN POTENTIAL ADVERSE EVENT ADDRESSED IN THE PRODUCT LABELING.
HEALTHCARE PROFESSIONAL REPORTED A LEFT SIDE RUPTURE. THE DEVICE HAS BEEN EXPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 708731 | STYLE 15 SILICONE GEL FILLED BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | ALLERGAN (COSTA RICA) | 2074755 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 YR | Required Intervention |