FDA Adverse Event Malfunction Summary report: N

ENDOPATH ENDOSCOPIC LINEAR CUTTER

MDR report key: 89068 · Received May 4, 1997

Report

Report Number
1527736-1997-00828
Event Type
Malfunction
Date Received
May 4, 1997
Report Date
April 29, 1997
Manufacturer
ETHICON ENDO-SURGERY, INC.
Product Code
GAG
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

H6; CODE 400: DAMAGED FIRING MECHANISM. FACILITY EXPERIENCED AN EVENT WITH THE ENDOPATH ENDOSCOPIC LINEAR CUTTER WHILE PERFORMING A LAP OOPHALERCTOMY. THE PRODUCT COMPLAINT ANALYSIS TEAM HAS COMPLETED ITS INVESTIGATION OF THE DEVICE WHICH WAS RETURNED TO CO WITH PRODUCT INQUIRY #972456-2. THE RESULTS OF THE INVESTIGATION CONDUCTED BY THE APPROPRIATE ENGINEERS AND TECHNICIANS ARE LISTED BELOW. VISUAL INSPECTIONS & RESULTS: LOCKOUT POSITION, BROKEN; CARTRIDGE CONDITION, PARTIALLY FIRED; CARTRIDGE RETURN BATCH NUMBER, H00181; AND INSTRUMENT NUMBER, 009. FUNCTIONAL TESTS & RESULTS: CONDITION OF FIRING TRIGGER LOCKOUT, CONDITION OF PINION GEAR, CONDITION OF SHORT RACK, AND CONDITION OF YOKE, GOOD; AND WAS INSTRUMENT CYCLED, NO. ANALYSIS CONCLUSION: BASED UPON THE INFO RECEIVED AND THE VISUAL EXAMINATION, IT WAS CONCLUDED THAT THE INSTRUMENT WAS RETURNED NON-FUNCTIONAL. THE INSTRUMENT WAS DISASSEMBLED AND FOUND TO HAVE A DAMAGED FIRING MECHANISM. NO CONCLUSION COULD BE REACHED AS TO HOW THIS DAMAGE OCCURRED. DUE TO THE DAMAGED FIRING MECHANISM THE INSTRUMENT MAY BECOME DIFFICULT TO RELEASE FROM TISSUE. SOME CONDITIONS WHICH MIGHT RESULT IN DAMAGE TO THE FIRING MECHANISM ARE: INTERRUPTED FIRING CYCLE, TISSUE THICKER THAN INDICATED, ATTEMPTING TO FIRE THROUGH A SPENT CARTRIDGE, FIRING PRIOR TO COMPLETE CLAMPING OF THE JAWS AND FAILURE TO PROPERLY FOLLOW REOADING INSTRUCTIONS. CO STRIVES TO UNDERSTAND EACH INCIDENT AS IT OCCURS IN ORDER TO CONTINUOUSLY IMPROVE THE PRODUCTS.

Description of Event or Problem · 1

THE INSTRUMENT WAS USED DURING A LAPAROSCOPIC OVARIAN CYSTECTOMY. IT WAS REPORTED THE INSTRUMENTS WOULD NOT FIRE PROPERLY, A CRUNCHING SOUND WAS HEARD, AND THERE WAS DIFFICULTY RELEASING THE ANVIL FROM THE TISSUE. THE CASE WAS COMPLETED WITHOUT THE EZ35. THERE WAS NO CONSEQUENCE TO THE PT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENDOPATH ENDOSCOPIC LINEAR CUTTER ENDOSCOPIC LINEAR CUTTER GAG ETHICON ENDO-SURGERY, INC. NA H41H6K

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN Other