ENDOPATH ENDOSCOPIC LINEAR CUTTER
Report
- Report Number
- 1527736-1997-00828
- Event Type
- Malfunction
- Date Received
- May 4, 1997
- Report Date
- April 29, 1997
- Manufacturer
- ETHICON ENDO-SURGERY, INC.
- Product Code
- GAG
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
H6; CODE 400: DAMAGED FIRING MECHANISM. FACILITY EXPERIENCED AN EVENT WITH THE ENDOPATH ENDOSCOPIC LINEAR CUTTER WHILE PERFORMING A LAP OOPHALERCTOMY. THE PRODUCT COMPLAINT ANALYSIS TEAM HAS COMPLETED ITS INVESTIGATION OF THE DEVICE WHICH WAS RETURNED TO CO WITH PRODUCT INQUIRY #972456-2. THE RESULTS OF THE INVESTIGATION CONDUCTED BY THE APPROPRIATE ENGINEERS AND TECHNICIANS ARE LISTED BELOW. VISUAL INSPECTIONS & RESULTS: LOCKOUT POSITION, BROKEN; CARTRIDGE CONDITION, PARTIALLY FIRED; CARTRIDGE RETURN BATCH NUMBER, H00181; AND INSTRUMENT NUMBER, 009. FUNCTIONAL TESTS & RESULTS: CONDITION OF FIRING TRIGGER LOCKOUT, CONDITION OF PINION GEAR, CONDITION OF SHORT RACK, AND CONDITION OF YOKE, GOOD; AND WAS INSTRUMENT CYCLED, NO. ANALYSIS CONCLUSION: BASED UPON THE INFO RECEIVED AND THE VISUAL EXAMINATION, IT WAS CONCLUDED THAT THE INSTRUMENT WAS RETURNED NON-FUNCTIONAL. THE INSTRUMENT WAS DISASSEMBLED AND FOUND TO HAVE A DAMAGED FIRING MECHANISM. NO CONCLUSION COULD BE REACHED AS TO HOW THIS DAMAGE OCCURRED. DUE TO THE DAMAGED FIRING MECHANISM THE INSTRUMENT MAY BECOME DIFFICULT TO RELEASE FROM TISSUE. SOME CONDITIONS WHICH MIGHT RESULT IN DAMAGE TO THE FIRING MECHANISM ARE: INTERRUPTED FIRING CYCLE, TISSUE THICKER THAN INDICATED, ATTEMPTING TO FIRE THROUGH A SPENT CARTRIDGE, FIRING PRIOR TO COMPLETE CLAMPING OF THE JAWS AND FAILURE TO PROPERLY FOLLOW REOADING INSTRUCTIONS. CO STRIVES TO UNDERSTAND EACH INCIDENT AS IT OCCURS IN ORDER TO CONTINUOUSLY IMPROVE THE PRODUCTS.
THE INSTRUMENT WAS USED DURING A LAPAROSCOPIC OVARIAN CYSTECTOMY. IT WAS REPORTED THE INSTRUMENTS WOULD NOT FIRE PROPERLY, A CRUNCHING SOUND WAS HEARD, AND THERE WAS DIFFICULTY RELEASING THE ANVIL FROM THE TISSUE. THE CASE WAS COMPLETED WITHOUT THE EZ35. THERE WAS NO CONSEQUENCE TO THE PT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENDOPATH ENDOSCOPIC LINEAR CUTTER | ENDOSCOPIC LINEAR CUTTER | GAG | ETHICON ENDO-SURGERY, INC. | NA | H41H6K |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN | Other |