SYR 10 ML FIL SALINE SHORTLENGTH PLUNGER ROD
Report
- Report Number
- 1911916-2019-00831
- Event Type
- Malfunction
- Date Received
- August 19, 2019
- Date of Event
- July 10, 2019
- Report Date
- August 9, 2019
- Manufacturer
- BD MEDICAL (BD WEST) MEDICAL SURGICAL
- Product Code
- NGT
- UDI-DI
- 00382903064991
- PMA / PMN Number
- K161552
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION SUMMARY: ONE SAMPLES WAS RECEIVED. IT HAS THE PLUNGER ROD- RUBBER STOPPER; NO TIP CAP AND NO SALINE SOLUTION. THE RUBBER STOPPER IS ALL THE WAY DOWN. THE PLUNGER ROD IS SEPARATED FROM THE RUBBER STOPPER. IT HAS BLOOD IN IT. THE POSIFLUSH PRODUCT IS DESIGNED TO FLUSH THE IV LINES, ONE SINGLE USE, NOT TO DRAWING BLOOD. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. INFORMATION WILL BE CAPTURED ON TREND REPORTS AND MONITORED MONTHLY. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS. A DEVICE HISTORY RECORD REVIEW WAS COMPLETED WITH ZERO DEFECTS FOUND. NO QUALITY NOTIFICATIONS WERE WRITTEN FOR THIS BATCH, NOR FOR THE ASSOCIATED ASSEMBLY BATCHES. INVESTIGATION CONCLUSION: THIS IS THE 1ST COMPLAINT FOR THE LOT# 9002663 FOR THE SAME DEFECT OR SYMPTOM. THERE WAS NO DOCUMENTATION OF ISSUES FOR THE COMPLAINT OF BATCH 9002663 DURING THE PRODUCTION RUN. ROOT CAUSE DESCRIPTION: ROOT CAUSE OFF LABEL USE. RATIONALE: THE POSIFLUSH PLATFORM WAS INFORMED OF THE OFF-LABEL USE TO ADDRESS THIS WITH THE CUSTOMER BY THE MARKETING TEAM.
IT WAS REPORTED THAT AN UNSPECIFIED NUMBER OF SYR 10 ML FIL SALINE SHORTLENGTH PLUNGER ROD EXPERIENCED DIFFICULT PLUNGER MOVEMENT AND STOPPER SEPARATION FROM PLUNGER DURING USE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL NO: 306499 , BATCH NO: 9002663. VERBATIM: THE FLUSHES ARE NOT COMING APART UNLESS LABS ARE BEING DRAWN AND THE NURSES ARE PULLING BACK TO OBTAIN THE WASTE. THE HANDLE CAN BE SCREWED BACK INTO THE RUBBER PIECE, BUT THAT IS NOT BEST PRACTICE EITHER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 698590 | SYR 10 ML FIL SALINE SHORTLENGTH PLUNGER ROD | INTRAVASCULAR CATHETER | NGT | BD MEDICAL (BD WEST) MEDICAL SURGICAL | 9002663 | 00382903064991 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 16 YR | Other |