PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAS
Report
- Report Number
- 2032227-2019-48652
- Event Type
- Injury
- Date Received
- August 19, 2019
- Date of Event
- August 7, 2019
- Report Date
- September 10, 2019
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- UDI-DI
- 00643169513914
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
DEVICE PASSED THE DISPLACEMENT TEST, REWIND, BASIC OCCLUSION, OCCLUSION, PRIME/A33 AND EXCESSIVE NO DELIVERY TEST. NO MOTOR ERROR ALARM NOTED. MOTOR PASSED MOTOR TEST. DEVICE RECEIVED WITH CRACKED RESERVOIR TUBE LIP. (B)(4).
(B)(4). CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY EXPERIENCED HYPERGLYCEMIA AND THE INSULIN PUMP RECEIVED MOTOR ERROR ALARM. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 435 MG/DL AND TREATED WITH MANUAL INJECTION. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING. CUSTOMER REPORTS THE DRIVE SUPPORT CAP WAS PROTRUDED. CUSTOMER WAS ABLE TO COMPLETE INSULIN PUMP REWIND. THE CUSTOMER ALSO REPORTED THAT THERE WERE MULTIPLE CRACKS ON TOP OF RESERVOIR COMPARTMENT. THE CUSTOMER WAS ADVISED TO DISCONTINUE THE USE OF INSULIN PUMP AND REVERT TO BACKUP PLAN. THE DEVICE WILL BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 699866 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-723LNAS | PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723LNAS | A5723LNASJ | 00643169513914 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 69 YR | Other | FRN-UNK-RSVR, UNOMED SET |