FDA Adverse Event
Malfunction
Summary report: N
THERMOPHORE
MDR report key: 8904852
·
Received August 19, 2019
Report
- Report Number
- 1832415-2019-10677
- Event Type
- Malfunction
- Date Received
- August 19, 2019
- Report Date
- July 30, 2019
- Manufacturer
- BATTLE CREEK EQUIPMENT
- Product Code
- IRT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
CUSTOMER STATED, "PAD WAS LAYING ON THE BED WHEN IT STARTED A FIRE." THE PRODUCT WAS RETURNED FOR INVESTIGATION. THE CUSTOMER DID NOT CLAIM ANY INJURY. THE PRODUCT WAS APPROX. 19 YEARS OLD WHEN THE INCIDENT OCCURRED. AN INVESTIGATION WAS DONE TO LOOK INTO THE CUSTOMERS COMPLAINT. INVESTIGATOR DETERMINED THE CUSTOMER WAS LAYING ON THE PAD DURING USE. THE LAYING ON THE PAD LED TO THE PVC TUBING ON THE HARNESS TO WEAKEN AND THE STAT INSIDE TO WEAKEN. THIS IS LIKELY THE CAUSE OF THE BURN THE CUSTOMER DESCRIBED. THE ALSO CUSTOMER ADMITTED TO LAYING ON THE UNIT. THE FAILURE THE CUSTOMER HAD WAS DUE TO THE CUSTOMER MISUSING THE PRODUCT. IFU STATES, "DO NOT SIT ON, LIE ON, OR CRUSH PAD. AVOID SHARP FOLDS".