FDA Adverse Event Malfunction Summary report: N

ADVANCED CONTROL BASE

MDR report key: 8903601 · Received August 16, 2019

Report

Report Number
2921578-2019-00032
Event Type
Malfunction
Date Received
August 16, 2019
Date of Event
August 12, 2019
Report Date
September 18, 2019
Manufacturer
MIZUHO ORTHOPEDIC SYSTEMS, INC.
Product Code
JEA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

DUE TO OPERATOR ERROR OF NOT STRAPPING PATIENT TO TABLE, THIS DIRECTLY CONTRIBUTED TO THE PATIENT FALLING OFF OF THE TABLE. THERE IS ALSO CONCERN WITH LACK OF SERVICE ON THE TABLE, AS MIZUHO OSI SERVICE RECORDS SHOW NO SERVICE ACTIVITY ON THIS TABLE SINCE IT WAS PURCHASED. ADDITIONALLY, THE SITE DOES NOT HAVE ANY TRAINED AND CERTIFIED PERSONNEL TO SERVICE THE TABLE, SO IT IS UNKNOWN WHAT THE SERVICE ACTIVITY HISTORY IS FOR THIS TABLE.

Additional Manufacturer Narrative · 0

PMA/510K: CORRECTED INFORMATION.

Description of Event or Problem · 0

IT WAS REPORTED WHEN THEY WERE TILTING THE TABLE, THE TABLE SLIPPED WHICH CAUSED THE PATIENT TO FALL OF THE TABLE. PATIENT WAS NOT STRAPPED TO THE TABLE TOP. UPDATE INFORMATION: PATIENT WAS NOT SECURED BY THE STRAP ON TABLE; MAY HAVE A POSSIBLE CONCUSSION FROM FALL TRANSFERRED TO ANOTHER HOSPITAL

Description of Event or Problem · 0

IT WAS REPORTED WHEN THEY WERE TILTING THE TABLE, THE TABLE SLIPPED WHICH CAUSED THE PATIENT TO FALL OF THE TABLE. PATIENT WAS NOT STRAPPED TO THE TABLE TOP. UPDATE INFORMATION: PATIENT WAS NOT SECURED BY THE STRAP ON TABLE; MAY HAVE A POSSIBLE CONCUSSION FROM FALL TRANSFERRED TO ANOTHER HOSPITAL.

Description of Event or Problem · 1

IT WAS REPORTED THAT MULTIPLE OCCURRENCES OF REDNESS/SKIN BREAKDOWN ON PATIENTS UNDERGOING PRONE SURGERY DURING MARCH 2019 TO JULY 2019. SECOND MDR 2921578-2019-00030 WAS CREATED TO REPORT THE SECOND PATIENT, AS THIS EVENT OCCURRED ON AT LEAST 2 PATIENTS. NO INFORMATION ABOUT THE SEVERITY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
696058 ADVANCED CONTROL BASE SURGICAL OPERATING TABLE JEA MIZUHO ORTHOPEDIC SYSTEMS, INC. 5803

Patients

Seq Age Sex Outcome Treatment
1 Other