BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET WITH PRE-ATTACHED HOLDER
Report
- Report Number
- 1024879-2019-01465
- Event Type
- Malfunction
- Date Received
- August 16, 2019
- Date of Event
- June 26, 2019
- Report Date
- September 24, 2019
- Manufacturer
- BECTON, DICKINSON & CO., (BD)
- Product Code
- FMI
- UDI-DI
- 50382903686563
- PMA / PMN Number
- K011984
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
INVESTIGATION: BD HAD NOT RECEIVED SAMPLES OR PHOTOS FROM THE CUSTOMER FACILITY FOR EVALUATION; THEREFORE, THE INVESTIGATION WAS LIMITED. A REVIEW OF THE DEVICE HISTORY RECORD WAS COMPLETED FOR THE INCIDENT LOT AND, BASED ON THIS REVIEW, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET; THERE WERE NO RELATED QUALITY NON-CONFORMANCES DURING MANUFACTURING OF THE PRODUCT. BD HAS INITIATED FURTHER INVESTIGATION RELATING TO THIS ISSUE THROUGH CAPA #982194 AND POTENTIAL CAUSES HAVE BEEN IDENTIFIED. AS A RESULT, CORRECTIVE ACTIONS HAVE BEEN ESTABLISHED AND ARE IN THE PROCESS OF BEING IMPLEMENTED.
IT WAS REPORTED THAT SEPARATION OCCURRED AFTER USE WITH A BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET WITH PRE-ATTACHED HOLDER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "BUTTERFLY NEEDLE GAUGE 23 - NEEDLE IS COMING APART BEFORE AND AFTER USE".
THERE WERE MULTIPLE LOT NUMBERS REPORTED TO BE INVOLVED. THE INFORMATION FOR EACH LOT NUMBER IS AS FOLLOWS: MEDICAL DEVICE LOT #: 9086811. MEDICAL DEVICE EXPIRATION DATE: 2021-03-31. DEVICE MANUFACTURE DATE: 2019-03-27. MEDICAL DEVICE LOT #: 8352537. MEDICAL DEVICE EXPIRATION DATE: 2020-12-31. DEVICE MANUFACTURE DATE: 2018-12-18. MEDICAL DEVICE LOT #: 9091626. MEDICAL DEVICE EXPIRATION DATE: 2021-03-31. DEVICE MANUFACTURE DATE: 2019-04-01. A DEVICE EVALUATION IS ANTICIPATED, BUT HAS NOT YET BEGUN. UPON COMPLETION OF THE INVESTIGATION, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT SEPARATION OCCURRED AFTER USE WITH A BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET WITH PRE-ATTACHED HOLDER. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER, "BUTTERFLY NEEDLE GAUGE 23 - NEEDLE IS COMING APART BEFORE AND AFTER USE."
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 697073 | BD VACUTAINER® PUSH BUTTON BLOOD COLLECTION SET WITH PRE-ATTACHED HOLDER | BLOOD SPECIMEN COLLECTION DEVICE | FMI | BECTON, DICKINSON & CO., (BD) | SEE H.10 | 50382903686563 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |